CaloSys Modules
51
Technical Data
Temperature range
18...36°C (64.4…96.8°F)
Resolution
0.01%
Accuracy
+/- 0.1°C (+/- 32.18°F)
Humidity
0...100% non-condensing
11.2.5.2. Digital Temperature Sensor
Technical Data
Operating range
-35°C…105°C
Resolution
12bit (corresponding to 0.0625°C)
Accuracy
+/- 0.5°C (-10°C…85°C)
Sensor
DS 18B20+
Connection
3-wire connection (data, GND, VDD) on 1-Wire bus
Insulating resistance >100MOhm, at 20°C (500V DC)
Protective conduit
1.4571 steel, V4A, thickness=6mm
Protection class
SELV / III (accord. to EN 60730)
Type of protection
IP65 / IP68 (accord. to EN 60529)
Conduit rolled, grouted, pressed
11.3. Calibration Master Process Control Unit
Allows to operate CaloSys module with automatic calibration.
Connections Calibration Master Process Control Unit
11.3.1.
System configuration
This section describes connections of a Calibration Master Control Unit of a CaloSys module without
Calibration Slave Control Units (see Sections 11.1.1.2 (push principle), 11.1.2.2 (pull principle)). For
instructions about connections of a Calorimetry Master Control Unit of CaloSys modules with
Calibration Slave Control Units, see Section 11.3.2. Unit IN and Gas OUT of the Calibration Master
Unit connect to the Calibration Slave Control Units (see Section 11.3.2). "XOext." connects to the
Calorimetry Master Control Unit. CaloSys module retrofitted for automatic calibration (hardware
upgrade): "X0ext." of Calibration Master Control Unit (Fig. 26) connects to the PC.
Fig. 24 Calibration Master Control Unit - Front panel
Содержание Calorimetry PhenoMaster
Страница 6: ...6 PhenoMaster Hardware Operating Instructions info TSE Systems com www TSE Systems com ...
Страница 13: ...CaloSys Modules 13 Confirm default installation path by pressing Next info TSE Systems com www TSE Systems com ...
Страница 15: ...CaloSys Modules 15 Press Finish info TSE Systems com www TSE Systems com ...
Страница 69: ...CaloSys Modules 69 info TSE Systems com www TSE Systems com ...