Calibration
Calibration adjustment is permanent and
should only be performed using accurate
calibration weights that match the
requirements for your scale (see the chart on
back). Do not use coins or other household
items to calibrate.
1.
Prepare the required weight for your scale. (see
chart on back)
2.
Find a flat and level surface to perform calibration
and let the scale adjust to room temperature.
3.
Press
ON/OFF
to turn the scale on and wait for the
scale to stabilize.
4.
Press and hold the
MODE
key until the display
shows ‘CAL’.
5.
Make sure nothing is on the platform then press
MODE
to confirm the zero point.
6.
The display will then flash the required calibration
weight.
7.
Place the required
calibration weight on the
center of the platform.
8.
The display will show ‘PASS’
then return to normal
weighing mode. You may
now turn the scale off or
continue weighing.