background image

Calibration

Calibration adjustment is permanent and 

should only be performed using accurate 

calibration weights that match the 

requirements for your scale (see the chart on 

back). Do not use coins or other household 

items to calibrate.

1. 

Prepare the required weight for your scale. (see 

chart on back)

2. 

Find a flat and level surface to perform calibration 

and let the scale adjust to room temperature.

3. 

Press 

ON/OFF

 to turn the scale on and wait for the 

scale to stabilize.

4. 

Press and hold the 

MODE

 key until the display 

shows ‘CAL’.

5. 

Make sure nothing is on the platform then press 

MODE

 to confirm the zero point.

6. 

The display will then flash the required calibration 

weight.

7. 

Place the required 

calibration weight on the 

center of the platform.

8. 

The display will show ‘PASS’ 

then return to normal 

weighing mode. You may 

now turn the scale off or 

continue weighing.

Radio Frequency

Interference

Unlevel Surfaces

High Humidity

Air Currents

Vibration

Extreme Temps

ON/OFF

PCS

MODE

TARE

Radio Frequency

Interference

Unlevel Surfaces

High Humidity

Air Currents

Vibration

Extreme Temps

ON/OFF

PCS

MODE

TARE

Radio Frequency

Interference

Unlevel Surfaces

High Humidity

Air Currents

Vibration

Extreme Temps

ON/OFF

PCS

MODE

TARE

All manuals and user guides at all-guides.com

Отзывы: