![True Wearables Oxxiom Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/true-wearables/oxxiom/oxxiom_user-manual_1175535046.webp)
true wearables
®
®
© 2016-2020 True Wearables, Inc.
UMD-0001 REV D
46
Warning delay features (seconds)
Warning description
Min
Typical Max
Oxxiom is not connected to iOS
device
1
2
3
Oxxiom’s battery is empty
1
2
3
Oxxiom is searching for valid
signal
8
(12)
12
(21)
15
(30)
Oxxiom wireless connection
with iOS device is not reliable
1
2
3
SpO2 and PR measurements
have crossed either upper or
lower limits
8
(12)
12
(21)
15
(30)
SpO2 measurement has
crossed either upper or lower
limits
8
(12)
12
(21)
15
(30)
PR measurement has crossed
either upper or lower limits
8
(12)
12
(21)
15
(30)
iOS device’s battery is low
1
2
3
NOTE: Delays in
Bold
are for measurement sites well perfused and not
subject to motion and/or signal interferences. Otherwise, the
corresponding delays in parentheses should be assumed.
OXXIOM WARNING SYSTEM, CONT.