4
airway features and is used for activation and
deactivation. For specific features, please see below;
Laryngospasm
1.
On the base’s interface, switch the laryngospasm control toggle from ‘off’ to
‘on.’ This opens up the system to allow air to travel to the mechanism.
2.
Pull o
ut the pump’s hose and attach the hose’s spin luer lock to the
corresponding luer lock on the base.
3.
When connected, use the pump to activate the mechanism - 3 repetitions fills
the mechanism fully.
4.
Remove the hose and return it to its retracted position.
5.
With the hose retracted fully, the feature will remain activated as long as the
control togg
le remains in the ‘on’ position.
6.
When finished, switch the control toggle to ‘off,’ this will release the air from
the mechanism and it will return to its inactivated state.
Displaced Larynx
1.
Place the head in your desired position, for best results we recommend
placing a pillow under the head or rotating the head into the sniffing position.
2.
On the base’s interface, switch the Displaced Larynx control toggle from ‘off’
to ‘on.’
3.
Pull out the pump’s hose and attach the hose’s spin luer lock to the
corresponding luer lock on the base.