The Model 3440 gauge uses operator-specified
target values
to
determine the percent compaction after determining the density
of the test material. Target values include
Gmb (Marshall)
,
Proctor
, and
Gmm (Voidless density)
values.
The gauge uses the Marshall and/or voidless density values to
calculate the compaction level in the
Asphalt
mode. The
voidless
density
is the maximum theoretical density for the test material
as obtained in laboratory tests. When enabled, the voidless
density is used to determine the percent air voids.
The gauge uses the Proctor value to calculate the compaction
level when taking measurements in the
Soil
mode.
The
Target
function enables you to enter and store a new target
value, as well as enable or disable a target value stored in
memory.
To access the
Target
menu, press the
¢
TARGET
²
key. The gauge
displays:
-Target-
1. GMB(Marshall)
2. Proctor
3. Gmm(Voidless)
NOTE
The
¢
TARGET
²
key is active only when the
Ready
screen is displayed.
To edit a Marshall, Proctor, or voidless density value, press the
number key that corresponds to the target value.
Содержание RoadReader 3430
Страница 14: ...1 6 Figure 1 Model 3440 Gauge and Accessories...
Страница 19: ...Model 3440 2 3 Figure 2 Direct Transmission Geometry Figure 3 Backscatter Geometry...
Страница 20: ...2 4 Figure 4 Backscatter Surface Density Effects Top Layer Effect Curves...
Страница 22: ...2 6 Figure 5 Effect of Moisture on Depth of Measurement...
Страница 164: ...F 2 STANDARD COUNT LOG Gauge Serial Number _____________ Date MS DS Date MS DS...