Appendix A–6
NOTE
Use only a light grade (ISO VG 32, SAE 10) hydraulic
oil to fill the hydraulic reservoir.
Hydraulic power unit – Examine all hoses and fittings for
leaks.
Hydraulic hoses and braided sleeve – Examine the hydraulic
hoses and their braided sleeves for wear. If the sleeve is
worn through and the rubber hydraulic hose is worn, replace
the hose and sleeve.
6
6
.
.
Open the chamber door and examine each hydraulic hose for
wear. If the rubber cover is worn, replace the hose.
EVERY 500 HOURS OF OPERATION OR ONCE A
YEAR
1
1
.
.
Remove the right-side service panel. Locate the hydraulic filter
(see Figure A–1). Verify that the bypass indicator on the top of
the filter housing is green. If the bypass indicator is red, the
filter element must be changed. Contact your Troxler
representative for more information.
Figure A
–1. Hydraulic Filter Bypass Indicator
ENSURE
BYPASS INDICATOR
IS GREEN
Содержание 5850v2
Страница 10: ...x LIST OF TABLES Table Title Page Table 2 1 Control Panel Keys and Button 2 2...
Страница 17: ...Model 5850v2 1 7 1 INTRODUCTION Figure 1 1 Model 5850v2 Gyratory Compactor...
Страница 27: ...Model 5850v2 2 3 2 SETUP OPERATION Figure 2 1 Model 5850v2 Control Panel Layout...
Страница 35: ...Model 5850v2 2 11 2 SETUP OPERATION part numbers 106952 or 106953...
Страница 37: ...Model 5850v2 2 13 2 SETUP OPERATION the chamber door...
Страница 62: ...4 8 Figure 4 3 Routing Load Cell Cables ROUTE CABLES THROUGH NOTCH LOAD CELL...
Страница 68: ...4 14 NOTES...
Страница 76: ...5 8 NOTES...
Страница 86: ...Appendix A 10 NOTES...
Страница 99: ...Model 5850v2 Index...
Страница 100: ...NOTES...
Страница 102: ...12...