Model 3450
5–5
STAT TEST
Erratic readings, or readings that seem to fluctuate, may indicate a
problem with the gauge. If the readings are suspect, perform a
stat
test
, or statistical stability test, to validate the normal operation of
the gauge.
A stat test consists of twenty 1-minute counts. From the twenty
counts, the gauge calculates the standard deviation. This standard
deviation is compared to a theoretical standard deviation value.
Ideally, this ratio should be 0.25, with acceptable limits from 0.17 to
0.33. The gauge is considered unstable if the ratio is outside these
limits and the stat test will fail. If the stat test fails, ensure that the
gauge setup and testing was conducted correctly, and take an
additional test. If the second test fails, contact the nearest Troxler
service center for assistance. For a list of Troxler and authorized
Troxler service centers, refer to page iii of this manual or visit the
Troxler website at www.troxlerlabs.com/SERVICES/services.shtml.
To access the
Stat Test
function, press
〈
4
〉
at the
Setup
menu. The
gauge displays the
Stat Test
menu:
To select a menu option, press the corresponding numeric key or
press
〈
ESC
〉
to return to the
Setup
menu.
TAKING A STAT TEST
Choose an area free of other nuclear gauges or radioactive sources.
Place the gauge on the reference standard block as shown in
To take a new stat test, press
〈
1
〉
at the
Stat Test
menu shown
above. The gauge prompts the operator to place the gauge on the
reference standard block with the source rod in the
SAFE
(shielded)
position. Ensure that the gauge is properly positioned and press the
〈
ENTER/START
〉
key.
5. SET
UP M
EN
U
Содержание 3430 Plus
Страница 3: ...Model 3430 Plus 3440 Plus iii ...
Страница 7: ...Model 3430 Plus 3440 Plus vii NOTES ...
Страница 10: ......
Страница 14: ...Model 3430 Plus 3440 Plus xii NOTES ...
Страница 16: ...xiv NOTES ...
Страница 30: ...2 6 NOTES ...
Страница 64: ...4 NOTES ...
Страница 98: ...7 NOTES ...
Страница 122: ...9 NOTES ...
Страница 142: ...Appendix A 20 NOTES ...
Страница 167: ...Model 3430 Plus 3440 Plus Appendix C 21 Figure C 4 Base Assembly Part Number 110015_XXXX C TROUBLESHOOTING ...
Страница 170: ...Appendix C 24 Figure C 5 Scaler Assembly Part Numbers 110791 3430 Plus and 110876 3440 Plus Continued ...
Страница 178: ...Appendix D 2 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Страница 180: ...Appendix D 4 STANDARD COUNT LOG Gauge Serial Number Date MS DS Date MS DS ...
Страница 190: ...Appendix F 8 NOTES ...
Страница 194: ...Appendix G 4 NOTES ...
Страница 200: ...Appendix H 6 NOTES ...
Страница 215: ...Model 3430 Plus 3440 Plus Index 11 W Wide Area Augmentation System WAAS 2 Y YES key 3 4 INDEX ...
Страница 216: ...Index 12 NOTES ...
Страница 217: ...Model 3430 Plus 3440 Plus Warranty 1 NOTES WARRANTY ...