Model 3400-B Series
3411-B Error Codes
Error
Code
Indicated Failure
Mode
Probable Cause
01
System failed internal
test condition.
Microprocessor memory board
hardware failure.
02
Accumulated number
exceeds display size.
Position of PWR/TIME switch
was changed during
accumulation OR hardware
failure on I/O board.
10,
11
Bad keyboard input.
Two keys depressed, defective
keyboard or hardware failure
on I/O board.
30
Microprocessor tried to
evaluate the log of a
negative number.
Incorrect depth selected by
DEPTH switch.
30
Microprocessor
attempted to divide by
zero.
No standard counts in the MS
or DS registers.
31
Negative overflow in
division.
Incorrect depth selected by
DEPTH switch OR standard or
measure counts are not valid.
32
Positive overflow in
division.
Same as above.
40
Invalid input from
MOISTURE
CORRECTION switches.
Switch failure, I/O board
failure, or operator induced
error during chain calculations.
41
Invalid input from
DEPTH switch.
Same as above.
Содержание 3411-B
Страница 3: ...Model 3400 B Series iii...
Страница 10: ......
Страница 15: ...Model 3400 B Series 1 3 Figure 1 Gauge and Accessories...
Страница 20: ...2 4 Figure 2 3411 B Front Panel Controls...
Страница 30: ...3 8 NOTES...
Страница 44: ...5 8 Figure 4 Roller Pattern...
Страница 56: ...6 6 The following is an example of the stat test and drift test Figure 9 Statistical Stability Test...
Страница 76: ...Appendix A 16 NOTES...
Страница 91: ...Model 3400 B Series Appendix C 3 Figure 15 Direct Transmission and Backscatter Geometries...
Страница 92: ...Appendix C 4 Figure 16 Backscatter Surface Density Effects...
Страница 104: ......
Страница 105: ...Model 3400 B Series Appendix E 1 A Ap pp pe en nd di ix x E E S St ta an nd da ar rd d C Co ou un nt t L Lo og g...
Страница 106: ...Appendix E 2 STANDARD COUNT LOG Serial Number Date Moisture MS Density DS Date Moisture MS Density DS...
Страница 108: ...Appendix E 4 NOTES...