Trotec BE49 Скачать руководство пользователя страница 1

TRT-BA-BE49-TC220328TTRT02-001-EN

BE49

EN

OPERATING MANUAL

DIGITAL MULTIMETER

Содержание BE49

Страница 1: ...TRT BA BE49 TC220328TTRT02 001 EN BE49 EN OPERATING MANUAL DIGITAL MULTIMETER...

Страница 2: ...an download the current version of the instructions and the EU declaration of conformity via the following link BE49 https hub trotec com id 46447 Safety Read this manual carefully before starting or...

Страница 3: ...ical data To use the device for its intended use only use accessories and spare parts which have been approved by Trotec Foreseeable misuse Do not use the device in potentially explosive atmospheres w...

Страница 4: ...ds on the position within a low voltage network in which a device machine is operated The closer this position is to the supply line the higher is the surge voltage to be expected This means that an e...

Страница 5: ...lay 2 HOLD button 3 Illumination button 4 Rotary switch 5 mA V socket 6 COM socket 7 Red measuring tip 8 Black measuring tip 9 Fold out stand 10 Fuse compartment below stand 11 Battery compartment 12...

Страница 6: ...170 g Dimensions length x width x height 121 mm x 65 mm x 35 mm Automatic switch off after 15 minutes of inactivity Safety This measuring device is designed for indoor use and complies with over volta...

Страница 7: ...ries before first use Note Disconnect the measuring tips from the device before opening the battery compartment Note Make sure that the surface of the device is dry and the device is switched off 1 Lo...

Страница 8: ...asuring device is not set to the current measuring range If the device indicates an exceedance of the measuring range OL immediately after being connected to the measuring object first switch off the...

Страница 9: ...cuit to be checked and connect the measuring device in series with the consumer in this circuit 2 Depending on the expected measuring current turn the rotary switch to position or 3 Use the MODE butto...

Страница 10: ...internal resistors of the measuring tips and sockets might lead to a falsified display The resistance value displayed in case of short circuited measuring tips will be put down in writing and later su...

Страница 11: ...With the display illuminated press the illumination button 3 to switch off the background illumination Switching the device off Info After 15 minutes of inactivity the device switches off automatical...

Страница 12: ...from Directive 2012 19 EU It states that this device must not be disposed of with the household waste at the end of its life You will find collection points for free return of waste electrical and el...

Страница 13: ...Trotec GmbH Grebbener Str 7 D 52525 Heinsberg 49 2452 962 400 49 2452 962 200 info trotec com www trotec com...

Отзывы: