pyrometer TP7

Technical data

Parameter

Value

Model

TP7

Weight

224 g

Dimensions
(length x width x height)

160 mm x 49 mm x 122 mm

Measuring range

-50 °C to 1000 °C (-58 °F to
1832 °F)

Resolution

0.1 °C / °F

Target display

Laser class II, 630 to 670 nm
< 1 mW

Accuracy

±2.5 °C (±4.5 °F) at -50 °C to
20 °C (-58 °F to 68 °F)
±1 % at 21 °C to 300 °C
(69 °F to 572 °F)
±1.5 % at 301 °C to 1000 °C
(573 °F to 1832 °F)

Degree of emission

Adjustable from 0.10 to 1.0

Optical resolution

40:1 (D:S)

Smallest measuring spot

ø 25.4 mm

Spectral sensitivity

8~14 μm

Response time

< 150 ms

Operating temperature

0 °C to 50 °C (32 °F to
122 °F), 10 % to 90 % RH

Storage conditions

-10 °C to 60 °C, < 80 % RH

Power supply

9 V battery

Switch-off

After approx. 10 seconds of
non-use

Scope of delivery

1 x Device 

TP7

1 x 9 V battery

1 x Quick guide

1 x Storage bag

Transport and storage

Note

If you store or transport the device improperly, the
device may be damaged.
Note the information regarding transport and storage of
the device.

Transport

For transporting the device use a suitable bag to protect it from
external influences.
The manufacturer packed the device to the best of his abilities
to protect it against transport damage.

Storage

When the device is not being used, observe the following
storage conditions:

dry and protected from frost and heat

protected from dust and direct sunlight

the storage temperature complies with the values specified
in the Technical data

Remove the battery from the device.

Operation

Inserting the battery

Note

Make sure that the surface of the device is dry and the
device is switched off.

21

22

1. Open the battery compartment by folding open the

cover (22) with your fingers.

2. Use the battery clip (21) to connect the new battery 

(1 x 9 V) with correct polarity.

3. Close the cover of the battery compartment.

Содержание 3510003012

Страница 1: ...TRT BA TP7 TC220613TTRT01 004 EN TP7 EN OPERATING MANUAL PYROMETER ...

Страница 2: ...s store the manual in the immediate vicinity of the device or its site of use Warning Read all safety warnings and all instructions Failure to follow the warnings and instructions may result in electric shock fire and or serious injury Save all warnings and instructions for future reference Do not use the device in potentially explosive rooms or areas and do not install it there Do not use the dev...

Страница 3: ... the Safety chapter Safety signs and labels on the device Note Do not remove any safety signs stickers or labels from the device Keep all safety signs stickers and labels in legible condition The following safety signs and labels are attached to the device Warning sign Meaning The warning sign is located on the back of the device and indicates that the device is equipped with a class 2 laser The p...

Страница 4: ... measurement are the measuring spot diameter and the degree of emission Measuring spot Observe the ratio of the distance D to the measuring spot diameter S The larger the distance to the object the larger the measuring spot diameter and hence the less precise the measured result The device calculates an average temperature from all the temperatures in the measuring spot Ø 12 mm 480 mm Ø 48 mm 1920...

Страница 5: ...n not oxidized 0 2 Skin 0 98 Haynes alloy 0 3 to 0 8 Radiator enamel 0 95 Timber natural 0 90 to 0 95 Inconel electro polished 0 15 Inconel oxidized 0 70 to 0 95 Inconel sand blasted 0 3 to 0 6 Limestone 0 95 to 0 98 Carborundum 0 9 Ceramics 0 88 to 0 95 Gravel 0 95 Carbon graphite 0 70 to 0 85 Carbon not oxidized 0 8 to 0 9 Material Degree of emission Plastic non transparent 0 95 Copper oxidized ...

Страница 6: ...N HOLD MIN C F L H 9 10 11 12 13 14 15 16 17 18 19 20 No Designation 9 SCAN indication 10 HOLD indication 11 Laser indication 12 Indication for continuous measurement 13 Indication of the upper alarm threshold 14 Indication of the lower alarm threshold 15 Temperature unit indication 16 Battery status indication 17 Indication of the degree of emission 18 MAX temperature indication 19 MAX indication...

Страница 7: ...bag Transport and storage Note If you store or transport the device improperly the device may be damaged Note the information regarding transport and storage of the device Transport For transporting the device use a suitable bag to protect it from external influences The manufacturer packed the device to the best of his abilities to protect it against transport damage Storage When the device is no...

Страница 8: ...g for the temperature unit is C This setting can be manually changed to F at the device Please proceed as follows to change the temperature unit 1 Repeatedly press the MODE button 6 until the Temperature unit indication 15 flashes 2 Press the Laser button 7 or the Lamp button 5 to change the temperature unit at the device Switching the laser pointer on or off Ex works the laser pointer is switched...

Страница 9: ...d as follows to set the upper and lower alarm thresholds 1 Repeatedly press the MODE button 6 until the corresponding icon flashes in the Indication of the upper alarm threshold 13 Indication of the lower alarm threshold 14 ð The characters OFF appear on the upper measurement value display 20 2 Press the Laser button 7 or the Lamp button 5 ð The characters ON appear on the upper measurement value ...

Страница 10: ...ions that apply for many EU countries on the website https hub trotec com id 45090 Otherwise please contact an official recycling centre for electronic and electrical equipment authorised for your country The separate collection of waste electrical and electronic equipment aims to enable the re use recycling and other forms of recovery of waste equipment as well as to prevent negative effects for ...

Страница 11: ...Trotec GmbH Grebbener Str 7 D 52525 Heinsberg 49 2452 962 400 49 2452 962 200 info trotec com www trotec com ...

Отзывы: