41 P5644.1600 Rev A
BreatheLITE
is designed to return a ‘Pass / Fail’ result based on the compliance audit check results and
operational threshold for the applicable particulate sizes. This is displayed on completion of the check
and is detailed in the device log of each connected XD One.
Return of a ‘Pass’ result
On return of a Pass result, the particulate sensor is functioning as expected and normal monitoring
operation can resume.
Return of a ‘Fail’ result
On return of a Fail result, the following procedure should be followed.
1.
Run a sensor cleaning operation as highlighted in section 15.2.
2.
Repeat the Compliance Audit Check sequence and note the test result.
3.
On return of a Pass result, the particulate sensor is functioning as expected and normal monitoring
operation can resume.
4.
On return of a repeat Fail result, please contact Trolex direct to discuss support or servicing of the
XD One unit.