INSTALLATION & OPERATING DATA
8 CALIBRATION
continued
8.4
Calibrate SPAN
Apply test gas, at a flow rate of 0.2 to 1 l/min, for a few seconds
until the display value stabilises.
The Gas Sensing Module will automatically return to
the NORMAL measuring mode if NO pushbuttons are
operated within 50 seconds.
Press CALIBRATE SPAN (5 seconds) until the indicator is flashing
YELLOW (3 times).
(SPAN CALIBRATION will not occur if an insufficient
concentration of span gas is present).
SCROLL the display value UP/DOWN until it concurs with the
value of gas concentration marked on the test gas canister.
50 seconds after SCROLLING has ceased the gas sensing
module will automatically return to the NORMAL measuring
mode.
TEST GAS
Y
Y
R
G
R
Keep the test gas applied for a few seconds to verify
the response of the sensor.
Switch off the test gas supply.