INSTALLATION & OPERATING DATA
Plug-in Gas
Sensing Module
Cable Entry
Gas Concentration
Display
Earth
Terminal
Gas Entry Port
7 CONTROLS AND INDICATORS
TX6373 TOXIC GAS SENSOR/ TRANSMITTER
G
Calibration STATUS indicator
SPAN calibrate pushbutton
ZERO calibrate pushbutton
TX6373.84 TOXIC GAS SENSOR/ TRANSMITTER REMOTE GAS SENSING MODULE.
This version uses the same pre-calibrated gas sensing module as the TX6373. The
module is fitted into a robust metal housing which can be mounted at a
remote location where space is constricted or the operating conditions are
extremely harsh.
Connections:
2 metres, flexible cable in a flexible armoured conduit
(other lengths available to specification - max. of 10 m)
Protection Classification: Dust and waterproof to IP66
Housing Material:
Brass or Stainless steel
Maximum Cable Length: Extendible up to 250 m using 1.5 mm
2
cable conductors
DISCONNECTING THE REMOTE GAS SENSING MODULE
The connecting cable between the remote gas sensing module
and the Transmitter is normally supplied connected at both ends.
The cable can be disconnected via a plug and socket connector
inside the Transmitter housing for servicing or transportation.
Secure connector to
inside of the enclosure
with a double sided
adhesive foam pad.
(Trim off any excess
before fitting).
Double line
indicates red marker
on ribbon cable.
I
I
M
M
P
P
O
O
R
R
T
T
A
A
N
N
T
T
!
!
The Ribbon Cable
Assembly must lie flat
against the inside rear
of the enclosure &
must be connected to
the feed through in the
orientation shown.
View showing
orientation of
ribbon cable.
(All other parts
removed for clarity).
13/25
ISSUE N 06/15