at least fifteen pipe diameters from a pipe
reducer and fifty pipe diameters from valves.
If this is not possible then the installation
of a standard flow straightener will improve
performance.
Fit a corresponding threaded boss or flange
at the monitoring point and install the sensor,
ensuring a air tight seal.
Checkpoint
High pressure versions, above 2 bar, will be
supplied with a welded bush or flange and
will require dedicated process fittings.
Release the clamping ring on the mounting
bush or flange.
The centre of the flow path of the sensing
head should be positioned as shown.
Checkpoint
Ensure that pressurised systems have
been completely vented before installation
or removal of the sensor.
Position the sensing head with the smaller
opening facing the flow, within a rotational
deviation of no more than 12.5° from the
axis of flow. Tighten the clamping ring with
moderate force.
Avoid fitting the sensor at low points
in pipework structures to prevent the
sensing head from being affected by large
accumulations of moisture.
www.trolex.com
Содержание Sentro Vortex Wireless TX5952
Страница 1: ......
Страница 2: ......
Страница 7: ...TX5952 User Manual TX5952 UM EN 01a 7 1 4 Dimensions 1 4 1 Side Projecting Sensor www trolex com...
Страница 8: ...8 TX5952 UM EN 01a 1 4 2 Remote Mounted Sensor www trolex com...
Страница 12: ...12 TX5952 UM EN 01a 2 Certification Pending www trolex com...
Страница 13: ...TX5952 User Manual TX5952 UM EN 01a 13 Certification continued Pending www trolex com...
Страница 18: ...18 TX5952 UM EN 01a 4 Setup and Calibration 4 1 Controls and Indicators www trolex com...
Страница 19: ...TX5952 User Manual TX5952 UM EN 01a 19 www trolex com...
Страница 20: ...20 TX5952 UM EN 01a 4 2 Software Menus www trolex com...
Страница 56: ......
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ......