Principles of Reverse Osmosis / Identifying System Components
Feed Water Requirements
TYPICAL EXPECTED ION REJECTION
IN A TFC MEMBRANE TYPE:
TYPICAL EXPECTED ION REJECTION IN A TFC MEMBRANE TYPE
SODIUM
85-94 %
LEAD
95-98 %
SULFATE
96-98 %
ARSENIC
92-96 %
CALCIUM
94-98 %
MAGNESIUM
94-98 %
POTASSIUM
85-95 %
NICKEL
96-98 %
NITRATE
60-75 %
FLUORIDE
85-92 %
IRON
94-98 %
MANGANESE
94-98 %
ZINC
95-98 %
CADMIUM
95-98 %
MERCURY
95-98 %
BARIUM
95-98 %
SELENIUM
94-96 %
CYANIDE
84-92 %
PHOSPHATE
96-98 %
CHLORIDE
85-92 %
WATER SUPPLY REQUIREMENTS:
TYPICAL REJECTION OF A REVERSE OSMOSIS MEMBRANE
TEMPERATURE
8-27° C
HARDNESS
< 20 PPM
RESIDUAL CHLORINE
0 PPM
TDS
< 1000 PPM
pH
5 to 6
IRON
< 0.01
SILICA
< 10 PPM
MANGANESE
< 0.05 PPM
INLET PRESSURE
30 - 60 PSI
II.- SYSTEM COMPONENTS
1. PREFILTER PRESSURE GAUGE:
Glycerin filled pressure gauge
(
0 -100 psi
). Standard on all units and is located on the front of the
panel. This gauge measures the water pressure in PSI at the outlet
of the pre-filter, a minimum pressure of 30 PSI is recommended.
2. SYSTEM PRESSURE GAUGE:
Glycerin filled pressure gauge
(0 -300 psi). Standard on all units. Located in front of the panel and
indicates the pump pressure discharge. This gauge measures the
pressure of water in PSI which is applied through the membrane.
(
Maximum recommended operating pressure is 180 - 200 PSI,
depending on the type of membrane
).
The operating pressure of the system can be adjusted by
regulating these combined valves:
- Turning waste valve handle to the right the pressure increases;
turning the valve handle to the left pressure will decrease.
- By adjusting the recirculation valve located next to the waste
valve above.
- Turn right to increase pressure and close recirculation flow
or turn the handle to the left to decrease pressure and open
recirculation flow.
3. CONCENTRATE WATER FLOWMETER (
WASTE
):
Located on
the front of the panel, this meter is calibrated in gallons per
minute. Its purpose is to measure the amount of waste water that
the system is producing.
4. PRODUCT WATER FLOWMETER (
PERMEATE
):
Located on
the front panel, its purpose is to measure the amount of product
water in gallons per minute.
5. REVERSE OSMOSIS MODULE:
It consists of a membrane
housing and the membrane. The assembly is placed horizontal
on IXC- 850 model.
6. WASTE VALVE:
This valve is located on the front panel. Its
purpose is to restrict the flow of concentrated water leaving the
module, it creates the necessary pressure inside the module so
that water is forced into the membrane creating the Reverse
Osmosis effect.
Turning the valve knob will control the ratio of water discharged/
water produced.
-
Turning clockwise will increase pressure inside of the module and
reduce the amount of waste water.
- Turning counterclockwise will decrease pressure within the
module and increase the amount of waste water.
CAUTION! DO NOT CLOSE THIS VALVE COMPLETELY.