Operation 

  

 

 

 

Viewing the Last Measured Value 

In the off state, short press the trigger to turn on the thermometer, it will 
display the last measurement data that was held before the last shutdown. 

 

Temperature Measurement 

Aim the thermometer at the target, then pull and hold the trigger, the laser 
indicator on the LCD display shows that the laser is turned on and points to 
the target, the measured temperature value will be displayed on the LCD, 
and the maximum or minimum temperature of the target area can be 
measured by scanning up, down, left and right according to the laser spot 
indication.  
Note: The measurement distance must be determined by the size of the 
measured targ

et according to the “distance to spot size ratio (D:S)” and “field 

of view”. 

 

Hold 

After the trigger is released, the last measured value and the HOLD icon will 
remain on the LCD display. If there is no operation for 8 seconds, the 
thermometer will automatically shut down and save the last measured value. 

 

Unit Conversion 

In the power-on state, open the battery cover and short press the °C/°F 
button to switch the temperature unit to Celsius (°C) or Fahrenheit (°F). 

 

Distance to Spot Size Ratio (D:S) 

As the distance (D) from the target being measured increases, the spot size 
(S) on the measured area becomes larger. The relationship between the 
distance and the spot diameter is as shown in the figure below.  
The spot diameter represents 90% of the energy in the circle. The maximum 
D:S (10:1) will be obtained when the thermometer is 400mm (16’’) from the 
target resulting in a spot diameter of 40mm (1.6’’). 

 
 
 

Содержание IRT225

Страница 1: ...ations on your purchase of the IRT225 10 1 infrared thermometer The IRT225 can quickly and accurately determine the surface temperature by measuring the infrared energy radiated from the target surfac...

Страница 2: ...re using the thermometer Inspect the case before using the thermometer Do not use the thermometer if it appears damaged Look for cracks or missing plastic Please refer to the emissivity information fo...

Страница 3: ...Description METER DESCRIPTION 1 Laser 2 Infrared sensor 3 Trigger 4 Battery cover 5 LCD 6 C F switching button 7 Battery compartment LCD DESCRIPTION 1 Function Indicator...

Страница 4: ...t according to the distance to spot size ratio D S and field of view Hold After the trigger is released the last measured value and the HOLD icon will remain on the LCD display If there is no operatio...

Страница 5: ...t diameter The smaller the target is the closer the test distance should be please refer to D S for the spot diameter at different distances To obtain the best measurement result it is recommended tha...

Страница 6: ...red value is less than the minimum range Stop measuring Err appears when booting Exceeding the minimum or maximum operating ambient temperature Place the thermometer in a 0 C 50 C 32 F 122 F environme...

Страница 7: ...the instructions for field of view D S etc Maintenance Cleaning and storage 1 The meter should be wiped with a damp cloth and mild detergent when necessary Do not use solvents or abrasives 2 Store th...

Страница 8: ...d at 90 energy Emissivity 0 95 Response time 500ms Spectral response 5 5um 14um Resolution 0 1 C 0 1 F Repeatability 1 C or 1 0 whichever is greater 2 F or 1 0 whichever is greater Laser Single laser...

Страница 9: ...ship and material for a period of 1 one year from the date of purchase This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or pur...

Отзывы: