P60 Setting the Real-Time Clock
1.
When in Real-Time Clock mode, P60 and “rtC” display on the
LCD. Short press the H/
button to advance to P60 if
necessary.
2.
Short press the MXMN/AV/
button to advance to P61 and
set the time as 12-hour or 24-hour format. The current time
format will flash on the LCD.
3.
Short press the H/
or MEM/
button to change the format
between 12-hour and 24-hour.
4.
Short press the MXMN/AV/
button to save the setting and
advance.
5.
The year will flash in the secondary display. Short press the
H/
or MEM/
button to change the value.
6.
Short press the MXMN/AV/
button to save the setting
and advance to the month setup.
7.
Short press the H/
or MEM/
button to change the value.
8.
Short press the MXMN/AV/
button to save the setting and
advance to the next step.
9.
Repeat the above steps to complete the setup for the
month/day/hour/minute/second settings.
10.
Short press the MXMN/AV/
button to save the selection and
return to P60.
11.
Short press the
/ESC button to exit setup mode.
Calibration
RH CALIBRATION
The meter is calibrated to a standard 33% and 75% salt solution
bottle. Calibration must be performed in a controlled
temperature environment of 25°C.
CAUTION: Do not calibrate the humidity without the default calibration
salts. The meter will lose its current calibration data and will become
inoperable.