5. Calibration 

 

The me

ter is calibrated at standard 400ppm CO2 concentration in factory. It’s suggested to do either ABC or 

manual calibration regularly to maintain good accuracy. 

 

Press 

“ SET”

 + 

“+”

keys at the same time till “

CAL

” is flashing. Long pressing 

“SET”

 key to start the 

calibration. During the calibration, 

“CAL”

 will flash on display with a 300 seconds count down. After 5 mins 

count down, the display return to normal measuring mode to indicate the calibration is done. During the 
calibration procedure, long pressing 

“+”

 key can quit the calibration function. 

 

CAUTION: 

Do not calibrate the meter in the air with unknown CO2 level.  Otherwise, it will be taken as 

 

400 ppm and leads to inaccurate measurement. The manual calibration is suggested to be done outdoor with 
good ventilation, fresh air and sunny day where CO

2

 level is around 400ppm. Do not calibrate in rainy day 

because high humidity will affect the CO

2

 level in air. Do not calibrate in places crowded with people or close to 

where exist high CO

2

 such as ventilating outlets or fireplaces. 

 

ERROR CODE AND SOLUTIONS 

 

 

CODE 

PROBLEM 

SOLUTION 

E01 

CO

2

 sensor is damaged 

Send back to repair 

E02 

Measured reading is under the lower limit 

For CO

2

 E02: Re-calibrate. For other E02: leave the meter in 

room condition  for 30 mins. 

 

 

If it not works, send back to repair 

E03 

Measured reading is above the upper limit 

For CO

2

 E03: put the meter in fresh air for 5 mins. For other 

E03: leave the meter in room condition  for 30 mins. 

 

 

If it not works, send back to repair 

 

 

 

E17 

ABC mode of CO

2

 sensor is fail 

Send back to repair 

E31

.

34 

Temp

./

RH measuring circuit is in error 

Send back to repair 

 

 

Can’t power on 

 

Check whether the adaptor is well plugged and the 5VDC power supply is working. 

Warranty 

 

 
 
 
Triplett / Jewell Instruments extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the 
original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (1) one 
year from the date of purchase. This warranty does not apply to any of our products which have been repaired or altered by 
unauthorized persons in any way or purchased from unauthorized distributors so as, in our sole judgment, to injure their stability or 
reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers 
altered, defaced, or removed. Accessories, including batteries are not covered by this warranty 

Copyright © 2021 Triplett 

www.triplett.com 

 

Содержание GSM250

Страница 1: ...erature Wall mount display with light bar to provide easy reading from far distance Touch panel to program the setting NDIR non dispersive infrared waveguide CO2 technology 3 color light bar green yellow red gradually changes color with CO2 concentration CO2 Automatic background calibration called ABC for long time drift compensation Buzzer Alarm function Areas of application Widely used in the of...

Страница 2: ...through Not included in package USB type C USB type C CAUTION Please take off the AC power adapter and store it well when this CO2 monitor left idle for long Please use a standard USB power supply such as PC s USB port universal AC adapter with USB port Improper power supply can cause serious damage to the device or result in injury or death to the user The vents allow the air circulation liquid f...

Страница 3: ...2 50ppm 5 of reading 0 2000ppm Other range is 10 of reading Temperature 0 6oC 0 9oF 5 RH at 25o C 10 90 RH Humidity 7 RH at 25oC other range Response time CO2 30 seconds 63 step change Tair 20 mins 90 step change Humidity 20 minutes 90 step change Operating 14 140 F 10 60oC 0 95 RH avoid condensation o Power supply 5VDC USB type C port Power consumption 500mA Meter size 8 9 x 6 0 x 1 8 226 x 152 x...

Страница 4: ...t Buzzer ABC auto background calibration and bL Brightness of light bar Press key to adjust the parameters Note During the setting process the device will return to the main screen after 25 sec of idle time 4 1 Alarm AL The default alarm setting is 1 000ppm It is adjustable from 800ppm to 3 000ppm the interval is 100ppm This threshold is related to buzzer alarm trigger and red light bar trigger Pr...

Страница 5: ...g is under the lower limit For CO2 E02 Re calibrate For other E02 leave the meter in room condition for 30 mins If it not works send back to repair E03 Measured reading is above the upper limit For CO2 E03 put the meter in fresh air for 5 mins For other E03 leave the meter in room condition for 30 mins If it not works send back to repair E17 ABC mode of CO2 sensor is fail Send back to repair E31 3...

Отзывы: