![Trio CS-2100A Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/trio/cs-2100a/cs-2100a_instruction-manual_1163527007.webp)
SPECIFICATIONS
W E I G H T
7 4 kg
A C C E S S O R I E S
PC-29 Probes 2
Instruction M a n u a l 1
Hand Book 1
A C p o w e r cord 1
Panel Cover 1
Probe holder 1
O P T I O N
Accessory Bag ( M C - 7 8 )
PRECAUTIONS FOR USE
1) Do not apply input voltage exceeding their m a x i m u m
rating and . never apply external voltage to the input
terminals.
CH 1 and C H 2 :
1
Mil.
5 0 0 V p - p or 2 5 0 V (DC + A C peak)
500: 5 V r m s or DC ± 5 V
C H 3 and C H 4 : 5 0 V (DC •+- A C peak)
Z-axis Input: 5 0 V (DC + A C peak)
2) The f o l l o w i n g conditions should be avoided:
(a) Direct sunlight
(b) High temperatures and humidity
(c) M e c h a n i c a l vibration
(d) Proximity to electrical noise or transient generating
machinery.
3) Do not use more than t h e required b e a m intensity.
4) Do not leave the b e a m stationary for long periods.
5) The C S - 2 1 0 0 A makes use of internal battery w h i c h , if
d e p l e t e d , can cause the panel LED's n o t t o light upon tur-
ning t h e unit ON. If this occurs, refer to the instruction
manual for instructions on replacing the battery.
6) To prevent electrical shocks, be sure to c o n n e c t the G N D
terminal to an appropriate earth point.
7) Cooling precautions.
The C S - 2 1 0 0 A is provided w i t h a c o n v e n i e n t carrying han-
dle w h i c h doubles as a stand to adjust the v i e w i n g angle.
W h i l e any arbitrary angle m a y be set, be sure that no ob-
jects are a l l o w e d to rest on the t o p of the unit or that the
cooling vents are not blocked, since this w i l l cause an u n -
due t e m p e r a t u r e rise in the C S - 2 1 0 0 A .
8) Since the C S - 2 1 0 0 A makes use of high voltage circuitry, if
removing the case, refer t o the " M A I N T E N A N C E " for
r e m o v i n g the case.
9) AC voltage selector s w i t c h setting.
Stopper
M A D E I N J A P A N
M A D E I N J A P A N
1 Remove screw and stopper plate.
2 Switch lever to opposite side.
3 Lock lever by attaching stopper plate to opposite side screw.
2-3
E N V I R O N M E N T
Operating t e m p e r a t u r e and
h u m i d i t y for g u a r a n t e e d
specifications: 1 0 ~ 3 5 ° C , 8 5 % m a x i m u m RH
Full operating range: 0 ~ 5 0 ° C , 9 0 % m a x i m u m RH
Storage t e m p e r a t u r e and
humidity range: — 2 0 ° ~ 5 5 ° C
80% m a x i m u m
A l t i t u d e :
Operating: 5 0 0 0 m
N o n - o p e r a t i n g : 1 2 0 0 0 m
C A U T I O N :
Before applying p o w e r to the C S - 2 1 0 0 A be sure that the Line
Voltage S w i t c h is set to the proper voltage. If this is not done
d a m a g e to t h e C S - 2 1 0 0 A could result.
After verifying the setting, connect the A C P o w e r Cord to the
line p o w e r source.