Error Codes 

ERROR 

REASON 

SOLUTION 

UUUUUU 

1.    Overload 
2.  wrong  connection 
with load cell 
3. load cell has quality 
problem. 

1. reduce the weight 
2. check load cell connection 
3.  inspection  load  cell.  Check 
the input and output   

nnnnnnn 

1.  calibration  is  no 
good 
2. wrong connection 
3. load cell has quality 
problem 

1.  check  scale  is  resisted  or 

not, foot is kept level or not. 

2. check load cell connection. 
3.  checking  load  cell

check 

input  and  output  resistance  to 
judge it is good or not. 

ERR1 

 
During  calibration,  not 
input  the  weights  or 
the weight is overload 

Input the correct weights 

ERR2 

 
During calibration , the 
weights  is  below  than 
Min. required weights 

The 

calibration 

weights 

Minimum  is  10%  of  Max.  cap. 
Recommend 60%-80% of Max. 
Cap. 

ERR3 

 
During  calibration,  the 
input 

signal 

is 

negative 

1.  check  the  connection  is 

correct 

2.  check  load  cell  is  no 

problem 

3.  recalibration  if  still  wrong 

change the PCB 

ERR4 

During  calibration,  the 
signal is unstable 

After  the  platform  is  stable, 
start calibration 

ERR5 

 

Change PCB 

Содержание TS 700-SS

Страница 1: ...User s Guide TS 700 SS Digital Indicator Series 3 Version 3 0 11 15 10 Triner Scale Mfg Co Inc 800 238 0152...

Страница 2: ...unless done so by a qualified scale technician WARNING The indicator is the static and sensitive equipment cut off the power during electrical connections internal components touched by hand is prohi...

Страница 3: ...____ 3 Serial Port Connections __________________________ 4 Operation ______________________________________ 5 Parameters and Settings _________________________ 7 Data Output Formats _________________...

Страница 4: ...100 250V Operation temperature 10 C 40 C Operation humidity 90 RH Storage temperature 40 C 70 C Battery Operation Before operating with the internal battery charge the battery fully for 18 hours The b...

Страница 5: ...oad cells 4 wire or 6 wire configurations Excitation voltage for the load cell is 5VDC the largest output current 120mA Terminal Block on Circuit Board Circuit Board Terminals Load cell EXC voltage SI...

Страница 6: ...4 Serial Port Connection RS232 Terminal on Circuit Board DB9 Definition Function 2 TXD Sending data 3 RXD Receiving data 5 GND Ground interface Note if RS485 The connection pin is 2 and 5 pin...

Страница 7: ...Scale With weight on scale press key to hold weight To Hold Peak Weight of Multiple Weighings Without weight on scale press key to put indicator in peak hold mode To Remove Hold Function Press key aga...

Страница 8: ...If you wish to clear the totals memory press the Zerro Up Arrow key to change the Clr n display to Clr y Press the Print key to accept the Clear Yes prompt kg lb Change pounds kilograms Gross For use...

Страница 9: ...You can move forward through all parameters and view the settings by pressing the PRINT ENTER key repeatedly Navigate to the setting parameter you wish to adjust and use the arrow keys to change the s...

Страница 10: ...within 60 minutes Power saving setting C10 Options 0 close power saving setting 3 close display if no change within 3min 5 close display if no change within 5 min Hold function C11 Options 0 no hold f...

Страница 11: ...lay 2 Print method connect printer 3 Command request method connect computer 4 PC continues sending format connect computer 5 PC big display continuous sending format Communication setting C19 Baud ra...

Страница 12: ...manual zero range Zero tracking CAUTION Changing these parameters will effect the indicator s performance C23 Automatically zero tracking time Options 0 disable zero tracking time 1 1 second 2 2 secon...

Страница 13: ...ons 0 close dynamic filter 1 1 digital filter strength 2 2 digital filter strength 3 3 digital filter strength 4 4 digital filter strength 5 5 digital filter strength 6 6 digital filter strength CAUTI...

Страница 14: ...34 Communication add C34 0 99 Add Code RESERVED Wireless communication C35 C35 0 99 signal RESERVED Gravity of calibration location C36 CAUTION Changing these parameters will effect the indicator s pe...

Страница 15: ...6 State A Bits0 1 2 0 1 2 Decimal point position 1 0 0 XXXXXX0 0 1 0 XXXXXXX 1 1 0 XXXXX X 0 0 1 XXXX XX 1 0 1 XXX XXX Bits3 4 Division 0 1 X1 1 0 X2 State B BitsS function Bits0 gross 0 net 1 Bits1 S...

Страница 16: ...mode NT net mode S3 weight of positive and negative or S4 kg or lb Data weight value including decimal point CR carriage return LF line feed Serial Interface Reception Command RS232COM serial interfac...

Страница 17: ...eceive data format 5 4 Print format ID NO 004 Serial No Date XX XX XX yy mm dd Time XX XX XX hh mm ss GROSS 8 88kg gross weight TARE 2 88kg tare NET 6 00kg net weight 5 5 PC or Big display continuous...

Страница 18: ......

Страница 19: ...cking load cell check input and output resistance to judge it is good or not ERR1 During calibration not input the weights or the weight is overload Input the correct weights ERR2 During calibration t...

Страница 20: ...pacity calibration 0 C07 Restore default 0 C08 Warning tone 1 C09 Power off automatically 0 C10 Power saving mode 0 C11 Hold function 0 C12 Prohibit kg lb conversion C13 Upper limit alarm 000000 C14 U...

Страница 21: ...1 C27 Standstill range 2 C28 Dynamic filter 0 C29 Noisy filter 2 C30 Print format 0 C31 Analog signal options 1 C32 4 20mA testing 4 C33 Relay output setting 1 C34 Muti PC communication add 0 C35 Wire...

Отзывы: