8310 RSA Operation Manual
Return Sweep Analyzer
Setting the RX Level & Tap Value
To set the RX level and tap value, perform the following steps:
1. From the
SETUP MENU
, highlight
RX LEVEL
and then select the
Enter
button to display
the
RX LEVEL
menu.
2. Use the arrow buttons to highlight the
RX
Level
or
Tap Value,
and then select the
Enter
button to edit the level.
3. Use the up/down arrow buttons to adjust the
value of the selected digit, and then use the
left/right arrow buttons to change between
the digits. Once you have finished editing
the level, select the
Enter
button to accept
your changes. The 8310 RSA has a RX level
range of 0 to 60 dBmV and a tap value range
of 0 to 60 dB. The minimum tap level must be
within 15 dB of the current RX level, or the
instrument will automatically adjust the value
to bring the tap value within the proper range.
4. Use the arrow buttons to highlight
OK
and
then select the
Enter
button to save the RX
level and tap value.