
C-Series Product Manual
99
7
Troubleshooting
About this chapter
This chapter provides diagnostic and troubleshooting information for
the C-Series Precision Syringe Pumps.
The following parts are included:
• Performing diagnotic self test, page 99
• Symptoms and solutions, page 100
• Error code listing, page 101
Performing diagnotic self test
During the self test, the pump initializes, then cycles repeatedly through
a series of plunger movements. If an error condition occurs, the pump
stops moving.
To perform the self test:
1.
Turn power off.
2.
Turn the address switch to the F position.
3.
Install the AutoRun jumper on J2. For location of the AutoRun jumper,
see Figure 4-4 (page 23).
4.
Turn power on.
5.
If errors are experienced, refer to "Status and error codes" (page 85).
6.
Return address switch and AutoRun jumper to their normal position
when finished.
Содержание C24000
Страница 8: ...This page intentionally blank ...
Страница 14: ...This page intentionally blank ...
Страница 142: ...This page intentionally blank ...