FlexxPump4 BXXX-EXT
Supplementary Sheet
Revision: 1
13
EN
-
4. Remount the cover
Place the cover on the battery compart-
ment.
Turn the screws clockwise to fasten.
3. Inserting the battery packs
Connect the new battery to the battery
cable via the reverse polarity protected plug
or insert new batteries into the matching
battery holder and connect it to the reverse
polarity protected plug.
Stow the battery and the battery cable in
the battery compartment.
Make sure that the battery cable or the
plug does not protrude from the battery
compartment and become trapped when the
cover is screwed on.
Carry out the work according to the above description.
If an error was displayed in the LCD before the cartridge was changed, the indicator
now disappears, as does the red LED which flashes every 5 seconds. A separate ack
-
nowledgement is not required. The green LED flashes briefly every 60 seconds.
After completion of this work, the FP4 B automatically returns to the last active mode
(ON or OFF) before the work.
If the empty battery error (error E1) has been reached during a dispensing cycle, the
cycle will be automatically interrupted and automatically continued after completion of
the work.