Safety Information

The signs below might be in the user manual, labeling or other component. 

They are the requirement of standard and using.

Measurement Principle

   This product uses the Oscillometric Measuring method to detect blood pressure. 

Before every measurement, the unit establishes a “zero pressure” equivalent to the 

atmospheric pressure. Then it starts inflating the arm cuff, meanwhile, the unit detects 

pressure oscillations generated by beat-to-beat pulsatile, which is used to determine 

the systolic and diastolic pressure, and also pulse rate.

General Description

Indications for Use

The Transtek Blood Pressure Monitor is a digital monitor intended for use in 

measuring blood pressure and heartbeat rate with arm circumference ranging 

from 22 cm to 32 cm ( about 8¾˝-12½˝ ) or 22cm to 42cm(about 8¾˝-16½˝).

It is intended for adult indoor use only.

Thank you for selecting TRANSTEK arm type blood pressure monitor 

(TMB-1585-BS). The monitor features blood pressure measurement, pulse  

rate measurement and the result storage. The design provides you with 

two years of reliable service.

Readings taken by the TMB-1585-BS are equivalent to those obtained by  

a trained observer using the cuff and stethoscope auscultation method.

This manual contains important safety and care information, and  

provides step by step instructions for using the product.

Read the manual thoroughly before using the product.

Features:

 

Maximum 60 records per each user

3rd technonoly: Measuring during inflation 

100mm×68 mm Digital LCD display 

1.The device is not suitable for use on may be pregnant women or pregnant women.

2.The device is not suitable for use on patients with implanted,electrocical devices,

such as cardiac pacemakers, defibrillators.

Contraindications

2

3

INTRODUCTION

INTRODUCTION

Refer to instruction manual/booklet

To signify that the instruction manual/booklet must be read.

Symbol for “Manufacturer”

Symbol for “Serial Number”

Symbol for “Type BF applied part”

Symbol for “Direct Current”

The symbol indicates that the product should not be discarded 

as unsorted waste but must be sent to separate collection 

facilities for recovery and recycling.

Symbol for “Date and Country of manufacture”

For indoor use only

Symbol for “Class II Equipment”

SN

Symbol for “Recycle”

Caution

Indicates that caution is necessary when operating the device 

or control close to where the symbol is placed, or that the 

current situation needs operator awareness or operator action 

in order to avoid undesirable consequences

.

Содержание TMB-1585-BS

Страница 1: ...gshan Guangdong China Blood Pressure Monitor TMB 1585 BS Arm Type Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB 1585 BS Please read the user manual carefully and thoroughtly so...

Страница 2: ...NAGEMENT 16 Recall the Records Delete the Records INFORMATION FOR USER 19 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 21 What are systolic pressure and diastolic pressure What is the standa...

Страница 3: ...ned by a trained observer using the cuff and stethoscope auscultation method This manual contains important safety and care information and provides step by step instructions for using the product Rea...

Страница 4: ...high frequency HF surgical equipment at the same time CAUTION Warning This device is not AP APG equipment Do not use the device where flammable anesthetic are present or in environments mixture with...

Страница 5: ...nt Blood pressure level indicator Indicate the blood pressure level Blood pressure monitor is detecting a heartbeat during measurement Monitor Components List 1 Blood Pressure Monitor TMB 1585 BS 4 Us...

Страница 6: ...e shows The display is dim The display does not light up The Choice of Power Supply 1 Battery powered mode 6VDC 4 AAA batteries 2 AC adaptor powered mode 6V 1A Please only use the recommended AC adapt...

Страница 7: ...Repeat step 2 and 3 to set the HOUR and MINUTE 5 11 10 BEFORE YOU START BEFORE YOU START After the unit is set the LCD will display donE first then display all the settings you have done and then tur...

Страница 8: ...the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first measurement Wait at least 3 minutes between measurements This allow...

Страница 9: ...tance 10 meters 15 14 MEASUREMENT MEASUREMENT If the data transmission fails the LCD will display If the data transmission succeeds the with disappear 2 This device will proceed to data transmission a...

Страница 10: ...ing after measurement To enable the data transmission function this product should be paired to Bluetooth end at 2 4 GHz How to mitigate possible interference The range between the device and BT end s...

Страница 11: ...u did not get the correct measurement you can delete all results by following steps below Take User 1 for example 1 Hold pressing M button for 3 seconds when the monitor is in the memory recall mode t...

Страница 12: ...of 25 or there are four or more pulse intervals the difference between each interval and the average is more than the average value of 15 the irregular heartbeat symbol appears on the display when the...

Страница 13: ...uestions for problems you may encounter with your blood pressure monitor If the product is not operating as you think it should check here before arranging for servicing PROBLEM SYMPTOM CHECK THIS REM...

Страница 14: ...ation For more information about our products please visit www transtekcorp com you can get customer service usual problems and customer download transtek will serve you anytime Manufactured by Compan...

Страница 15: ...missions test This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any...

Страница 16: ...ngle phase at 0 0 UT 250 300 cycle Conduced RF IEC61000 4 6 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur radio bands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 3 V 0 15 MHz 80 MHz 6 V in ISM and amateur...

Отзывы: