Type BF applied part
WARNING: No modification of this equipment is allowed.
Power supply
Battery powered mode:
6VDC 4×AAA batteries
AC adaptor powered mode: 6V 1A
(Please only use the recommended AC adaptor
model).
Display mode
Measurement mode
Oscillographic testing mode
Measurement range
Measurement perimeter
of the upper arm
Weight
External dimensions
Attachment
Mode of operation
Continuous operation
Degree of protection
Protection against
ingress of water
Accuracy
Normal working condition
Storage & transportation
condition
Software Version
A01
Pressure:
5°C-40°C within±3mmHg(0.4kPa)
Pulse value:±5%
Rated cuff pressure:
0mmHg~299mmHg(0kPa ~ 39.9kPa)
Measurement pressure:
SYS: 60mmHg~230mmHg (8.0kPa~30.7kPa)
DIA: 40mmHg~130mmHg (5.3kPa~17.3kPa)
Pulse value: (40-199)beat/minute
Device Classification
Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor Powered Mode: Class II ME Equipment
Digital LCD V.A.100mm×68mm
Approx.154mm×106mm×57.1mm
Contact Information
For more information about our products, please visit www.transtekcorp.com. you
can get customer service, usual problems and customer download, transtek will serve
you anytime.
Manufactured by:
Company:
Address:
Authorized Component
1
.
please use the TRANSTEK
authorized adapter.
Guangdong Transtek Medical Electronics Co., Ltd.
Zone A, No.105 ,Dongli Road, Torch Development District,
528437 Zhongshan,Guangdong,China
Guangdong Transtek Medical Electronics Co., Ltd.
Temperature:-20°C to +60°C
A relative humidity range of ≤ 93%, non-condensing,
at a water vapour pressure up to 50hPa
BLJ06L060100P-U
100-240V 50-60Hz,0.2Amax
6V 1000mA
Adaptor
Type:
Input:
Output:
24
25
SPECIFICATIONS
AUTHORIZED COMPONENT
About 22cm~32cm or 22cm~42cm
Approx.282g(Excluding the dry cells and cuff)
4×AAA batteries,user manual,AC adapter
IP21 It means the device could protected
against solid foreign objects of 12.5mm and
greater, and protect against vertically falling
water drops.
A temperature range of :+5°C to +40°C
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour
partial pressure greater than 50 hPa
An atmospheric pressure range of :
700 hPa to 1060 hPa