Why does my blood pressure fluctuate 
throughout the day?

Is the result the same 
if measuring on the 
right arm?

Why do I get different 
blood pressure at home 
compared to the hospital?

1. Individual blood pressure varies 
multiple times everyday. It is also 
affected by the way you tie your cuff and 
your measurement position, so please 
take the measurement under the same 
conditions.
2. If the person takes medicine, the 
pressure will vary more.
3. Wait at least 3 minutes for another 
measurement.

The blood pressure is different even 
throughout the day due to weather, 
emotion, exercise etc. Also, there is 
the “white coat” effect, which means 
blood pressure usually increases in 
clinical settings.

It is ok for both arms, but there will 
be some different results for 
different people. We suggest you 
measure the same arm every time.

What you need to pay attention to 
when you measure your blood 
pressure at home: 

If the cuff is tied properly. 
If the cuff is too tight or too loose.
If the cuff is tied on the upper arm.
If you feel anxious.
Taking 2-3 deep breaths before 
beginning will be better for 
measuring.        
Advice: Relax yourself for 4-5 
minutes until you calm down.

18

19

ABOUT BLOOD PRESSURE

TROUBLE SHOOTING

PROBLEM

SYMPTOM

CHECK THIS

REMEDY

No power

Low Battery

Error

message

Display can 
not light up. 

Batteries are exhausted.

Replace with new batteries.

Install the batteries or plug in the 
adapter properly.

Replace with new batteries.

Batteries are installed 
incorrectly or Adapter is 
not plugged in properly.

Lo        shows  The battery is too low.

E 1 shows

E 3 shows

E 4 shows

The measurement 
failed.

The cuff is not wrapped
or wrapped incorrectly, 
or the cuff air plug is 
loose.

Refasten the cuff and insert
air tube plug correctly then
measure again.

Relax for 5 minutes

,

 and then

keep still, measure again.

Relax for 5 minutes and 
measure again.

Pulse is not detected
during measuring.

Loosen the clothing on the 
arm and then measure again. 

Warning
message

Relax for a moment. Refasten
the cuff and then measure again.
If the problem persists, contact
your physician.

out shows

Out of measurement
range

EExx shows 

Hardware error 
(xx can be some digital
symbol, such as 1, 2, 
etc.)

Turn off monitor and measure
again. If EEx still appears on 
the display, please contact the
retailer or our customer service.

E 5 shows

Failed to communicate 
with the server

Try a place with better signal, 
or contact customer service

E 6 shows

Radio communication 
failure

Contact customer service

Excessive body motion 
(such as shaking of the 
arm with the cuff on)
or Pulse is weak 
during a measuring.

E 2 or 
         shows

If any abnormality arises during use, please check the following points:

NOTE:

 

If the product still does not work, contact the Customer Service. Under no 

circumstance should you disassemble or attempt to repair the unit by yourself. 

The supply voltage is 
too high. 

Replace with the authorized 
adapter.

H bAt  shows

High Battery

Содержание LS802-GS

Страница 1: ...Logo Customized Logo Please read the user manual carefully and thoroughtly so as to ensure the safe usage of this product keep the manual well for further reference in case you have problems version 1...

Страница 2: ...INFORMATION FOR USER 14 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 16 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irregular heartbe...

Страница 3: ...ed electrical devices such as cardiac pacemakers defibrillators Contraindications This product uses the Oscillometric Measuring method to detect blood pressure Before every measurement the unit establ...

Страница 4: ...ition and storage state The typical service life is 10000 times Dispose of accessories detachable parts and the device according to the local guidelines This device is intended for indoor home use Thi...

Страница 5: ...rs when decteted during a measurement Refer to page 17 for more information Signal indica on Indicates the signal situation in the communication process Date Time display 1 2 3 4 5 6 7 9 Data transmis...

Страница 6: ...er to depart from the using utility power when you finish the measurement 6V 1A Do not use new and used batteries together Do not use different types of batteries together Do not dispose the batteries...

Страница 7: ...ft arm use your right arm Roll or push up your sleeve to expose the skin Make sure your sleeve is not too tight 2 3 5 6 Sit upright in a comfortable chair with your back against the backrest of the ch...

Страница 8: ...n is achieved Tip You can press the START STOP button to stop the measurement any time 1 When the monitor is off press the START STOP button to turn on and then it will complete the whole measurement...

Страница 9: ...e temperature environment Do not attempt to clean the reusable cuff with water and never immerse the cuff in water Tips for Measurement Within 1 hour after dinner or drinking Within 20 minutes after t...

Страница 10: ...P range Please contact a physician if your measuring result falls out of the range Please note that only a physician can tell whether your blood pressure value has reached a dangerous point Irregular...

Страница 11: ...Install the batteries or plug in the adapter properly Replace with new batteries Batteries are installed incorrectly or Adapter is not plugged in properly Lo shows The battery is too low E 1 shows E 3...

Страница 12: ...s equipment is allowed Display mode Measurement mode Oscillographic testing mode Measurement range Measurement perimeter of the upper arm Weight External dimensions Mode of operation Continuous operat...

Страница 13: ...ions flicker emissions IEC 61000 3 3 Emissions test FCC Statement This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause ha...

Страница 14: ...d manufacturer s declaration electromagnetic Immunity Immunity Test 8 kV contact 2 kV 4 kV 8 kV 15 kV air 8 kV contact 2 kV 4 kV 8 kV 15 kV air 1 kV 2 kV 100 kHz repetition frequency For AC power port...

Отзывы: