background image

    This section includes a list of error messages and frequently 

asked questions for problems you may encounter with your blood 

pressure monitor. If the products not operating as you think it 

should, check here before arranging for servicing.

Power supply

Battery powered mode: 
6VDC 4*AA alkaline batteries
AC adaptor powered mode:
100-240V~, 50-60HZ,400mA
(Can be supplied by AC adaptor model 
UE08WCP-060100SPA only!)

Display mode

Digital  LCD  V.A.78*92mm

Measurement mode

Oscillographic testing mode

Measurement range

Accuracy

Normal working condition

Storage & transportation

condition

Measurement perimeter

of the upper arm

About 22cm~32cm or 22cm~42cm

Net Weight

Approx.

385g

(Excluding the dry cells)

External dimensions

Attachment

Approx.120*160*69mm

4*AA alkaline batteries,user manual  

Mode of operation

Continuous operation

Degree of protection

Type BF applied part

Protection against 

ingress of water

IP21

Device Classification

Battery powered mode: Internally powered ME Equipment
AC adaptor powered mode: CLASS II ME Euipment

Software Version

V01

15

14

SPECIFICATIONS

TROUBLESHOOTING

PROBLEM SYMPTOM CHECK THIS

REMEDY

No power

Low

batteries

Error

massage

Display will not 
light up.

Batteries are exhausted.

Replace with new batteries

Insert the batteries 
correctly

Replace with new batteries

Batteries are inserted

incorrectly.

Display is dim

or show

Batteries are low.

E 1 shows

RF communication failed.

Check the database is 

power on or not. 

Or synchronize the data 

operation.

E 3 shows

E10 or E11

 shows

E20 shows

Exxx,shows on 

the display.

A calibration error

occurred.

Retake the measurement.

If the problem persists,

contact the retailer or our

customer service 

department for further 

assistance.Refer to the 

warranty for contact 

information and return 

instructions.

E21 shows

AC adaptor is inserted

incorrectly.

Insert the AC adaptor 
tightly

The cuff is not secure.

Refasten the cuff and then

measure again.

The monitor detected

motion,talking or the

pluse is too poor 

while measuring.

Relax for a 

moment and then 

measure again.

The measurement 

process does not detect

the pulse signal.

Loosen the clothing on the

arm and then measure 

again

The treatment of the

measurement failed.

Relax for a moment and

then measure again.

E 8 shows

Failure of activating or 

synchronizing the network

Check if the newwork is

connected normally or not.

E 9 shows

The device is not 

activated or registered.

Activate or register the

device again.

Rated cuff pressure: 

0kpa - 40kpa (0mmHg~300mmHg)

Measurement pressure: 5.3kPa-30.7kPa

(40mmHg-230mmHg)

pulse value: (40-199) beat/minute

Pressure:
5

-40

within±0.4kpa(3mmHg) 

pulse value:±5%

Temperature:5

 to 40

 Relative humidity ≤85%

Atmospheric pressure: 86kPa to 106kPa

Temperature:-20

-60

Relative Humidity 10%-93%

Atmospheric Pressure: 50-106 kPa

Содержание LS802-E

Страница 1: ...u very much for selecting TRANSTEK Blood Pressure Monitor LS802 E please keep well this manual in order to reference in future Arm Type Version 1 0 FCC ID OU9LS802 E02 GUANGDONG TRANSTEK MEDICAL ELECTRONICS CO LTD Zone A 5 F Investment Building No 12 Huizhan East Rd Torch Development District Zhongshan Guangdong 528437 China ...

Страница 2: ...SER 10 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 12 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Why my blood pressure is varies even in one day Why the blood pressure I get from the hospital is different from home If the result is the same if measuring on the right arm TROUBLESHOOTING 14 SPECIFICATIONS 15 AUTHORIZED COMPONENTS 1...

Страница 3: ...e START STOP button to stop inflation The equipment is not AP APG equipment and not suitable for use in the presence of a flammable anesthetic mixture with air of with oxygen or nitrous oxide The operator shall not touch output of batteries adapter and the patient simultaneously To avoid measurement errors please avoid the condition of strong electromagnetic field radiated interference signal or e...

Страница 4: ...measurement for user 2 and then save the measure data Data storage To remind the users that the measurement data don t upload to bridge in time Wireless transmitter The blood monitor and Bridge in communication Network connection The bridge not connect the network Shocking remainding Shocking will result in inaccurate Low battery Batteries are low and need to be replaced Measurement Unit of the bl...

Страница 5: ...ws your blood circulation to recover For a meaningful comparison try to measure under similar conditions For example take daily measurements at approximately the same time on the same arm or as directed by a physician Tie the Cuff 1 Tie the cuff on your upper arm the position the tube off center toward the inner side of arm in line with the little finger 2 The cuff should be sung but not too tight...

Страница 6: ...ssful activation synchronization the equipment will display and transmit the measuring results after each measurement Activation synchroniz ation is proceeding The measureme nt data is uploading The data has been uploaded successfully CAUTION Interference may occur in the vicinity of equipment marked with the following symbol And LS802 E may interfering vicinity electrical equipment Sensitive peop...

Страница 7: ... it with a dry cloth in case Avoid the dusty and unstable temperature environment Avoid washing the cuff 11 10 Tips for measurement In a very cold environment Immediate measurement after tea coffee smoking When talking or moving your fingers When you want to discharge urine It can cause incorrectness if the measurement are taken in the following circumstances Within 20 minutes after taking a bath ...

Страница 8: ...LOOD PRESSURE If the result is the same if measuring on the right arm ABOUT BLOOD PRESSURE Irregular Heartbeat Detector This Blood Pressure Monitor is equipped with an intelligent function of Irregular Heartbeat IHB Detector During each measurement this equipment records the heartbeat intervals and works out the standard deviation If the calculated value is larger than or equal to 15 this equipmen...

Страница 9: ...th new batteries Batteries are inserted incorrectly Display is dim or show Batteries are low E 1 shows RF communication failed Check the database is power on or not Or synchronize the data operation E 3 shows E10 or E11 shows E20 shows Exxx shows on the display A calibration error occurred Retake the measurement If the problem persists contact the retailer or our customer service department for fu...

Страница 10: ...This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harm...

Страница 11: ...cle processes ISO EN 14971 2012 Medical devices Application of risk management to medical devices ISO EN 15223 1 2012 Medical devices Symbols to be used with medical device labels labelling and information to be supplied General requirements EN 1041 2008 Medical equipment manufacturers to provide information IEC 60601 1 2005 A1 2012 Medical electrical equipment Part 1 General requirements for basi...

Отзывы: