This section includes a list of error messages and frequently
asked questions for problems you may encounter with your blood
pressure monitor. If the products not operating as you think it
should, check here before arranging for servicing.
Power supply
Battery powered mode:
6VDC 4*AA alkaline batteries
AC adaptor powered mode:
100-240V~, 50-60HZ,400mA
(Can be supplied by AC adaptor model
UE08WCP-060100SPA only!)
Display mode
Digital LCD V.A.78*92mm
Measurement mode
Oscillographic testing mode
Measurement range
Accuracy
Normal working condition
Storage & transportation
condition
Measurement perimeter
of the upper arm
About 22cm~32cm or 22cm~42cm
Net Weight
Approx.
385g
(Excluding the dry cells)
External dimensions
Attachment
Approx.120*160*69mm
4*AA alkaline batteries,user manual
Mode of operation
Continuous operation
Degree of protection
Type BF applied part
Protection against
ingress of water
IP21
Device Classification
Battery powered mode: Internally powered ME Equipment
AC adaptor powered mode: CLASS II ME Euipment
Software Version
V01
15
14
SPECIFICATIONS
TROUBLESHOOTING
PROBLEM SYMPTOM CHECK THIS
REMEDY
No power
Low
batteries
Error
massage
Display will not
light up.
Batteries are exhausted.
Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Display is dim
or show
Batteries are low.
E 1 shows
RF communication failed.
Check the database is
power on or not.
Or synchronize the data
operation.
E 3 shows
E10 or E11
shows
E20 shows
Exxx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
E21 shows
AC adaptor is inserted
incorrectly.
Insert the AC adaptor
tightly
The cuff is not secure.
Refasten the cuff and then
measure again.
The monitor detected
motion,talking or the
pluse is too poor
while measuring.
Relax for a
moment and then
measure again.
The measurement
process does not detect
the pulse signal.
Loosen the clothing on the
arm and then measure
again
The treatment of the
measurement failed.
Relax for a moment and
then measure again.
E 8 shows
Failure of activating or
synchronizing the network
Check if the newwork is
connected normally or not.
E 9 shows
The device is not
activated or registered.
Activate or register the
device again.
Rated cuff pressure:
0kpa - 40kpa (0mmHg~300mmHg)
Measurement pressure: 5.3kPa-30.7kPa
(40mmHg-230mmHg)
pulse value: (40-199) beat/minute
Pressure:
5
℃
-40
℃
within±0.4kpa(3mmHg)
pulse value:±5%
Temperature:5
℃
to 40
℃
Relative humidity ≤85%
Atmospheric pressure: 86kPa to 106kPa
Temperature:-20
℃
-60
℃
Relative Humidity 10%-93%
Atmospheric Pressure: 50-106 kPa