Transmille EA002 Скачать руководство пользователя страница 12

EA002 2/10/50 Turn Clamp Coil Adapter Specifications

General Specifications

Adapter Connection

4x 4mm safety sockets mounted on the rear of the unit

Adapter Dimensions

28cm x 12cm x 6cm

Colour

Cream

Connection to Calibrator

Via supplied 4mm to 4mm connection leads (x2) to 3000 series current terminals

Coil Configuration

2 Turn (LHS) : 10 Turn (RHS) : 50 Turn (CENTRE)

Coil Type

High accuracy balanced configuration

Min. internal jaw dimensions

10mm (2 Turn & 10 Turn) : 25mm (50 Turn)

Maximum Current

30A

Maximum RMS voltage

4V

Frequency Range

DC to 500Hz

Construction

Loose wound coil (for heat dissipation) in moulded ABS enclosure

Durability

Fully enclosed coil for maximum protection from mechanical damage

Compatibility

Designed for use with Transmille 3000 Series calibrators and ProCal Software

2 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 60A)

90 Day Rel.

180 Day Rel.

1 Year Rel.

2 Year Rel.

%

A

%

A

%

A

%

A

Effective accuracy - Coil only (wound clamps)

0.35 + 0.008 0.35 + 0.008 0.35 + 0.008 0.35 + 0.008

Effective accuracy - Coil only (hall effect clamps)

0.48 + 0.07 0.48 + 0.07 0.48 + 0.07 0.48 +

0.07

Total uncertainty with 3050 (All clamps)

0.51 + 0.09 0.51 + 0.09 0.52 + 0.09 0.56 +

0.10

Total uncertainty with 3041 (wound clamps)

0.36 + 0.028 0.36 + 0.028 0.36 + 0.028 0.38 + 0.036

Total uncertainty with 3041 (hall effect clamps)

0.49 + 0.090 0.49 + 0.090 0.49 + 0.090 0.50 + 0.098

Total uncertainty with 3010 calibrator (wound clamps)

0.36 + 0.010 0.36 + 0.010 0.36 + 0.010 0.37 + 0.010

Total uncertainty with 3010 (hall effect clamps)

0.48 + 0.072 0.49 + 0.072 0.49 + 0.072 0.49 + 0.072

10 Turn Coil Accuracy (Input 0 to 30A : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 300A)

90 Day Rel.

180 Day Rel.

1 Year Rel.

2 Year Rel.

%

A

%

A

%

A

%

A

Effective accuracy - Coil only (wound clamps)

0.41 + 0.01 0.41 + 0.01 0.41 + 0.01 0.41 +

0.01

Effective accuracy - Coil only (hall effect clamps)

0.59 + 0.11 0.59 + 0.11 0.59 + 0.11 0.59 +

0.11

Total uncertainty with 3050 (All clamps)

0.61 + 0.13 0.62 + 0.13 0.62 + 0.13 0.65 +

0.14

Total uncertainty with 3041 (wound clamps)

0.42 + 0.03 0.42 + 0.03 0.42 + 0.03 0.43 +

0.04

Total uncertainty with 3041 (hall effect clamps)

0.60 + 0.13 0.60 + 0.13 0.60 + 0.13 0.61 +

0.14

Total uncertainty with 3010 calibrator (wound clamps)

0.41 + 0.012 0.42 + 0.012 0.42 + 0.012 0.42 + 0.012

Total uncertainty with 3010 (hall effect clamps)

0.59 + 0.112 0.59 + 0.112 0.60 + 0.112 0.60 + 0.112

50 Turn Coil Accuracy (Input 0 to 30A  : Freq. DC - 30Hz to 60Hz : Effective Output 0 to 1500A)

90 Day Rel.

180 Day Rel.

1 Year Rel.

2 Year Rel.

%

A

%

A

%

A

%

A

Effective accuracy - Coil only (wound clamps)

0.24 + 0.04 0.24 + 0.04 0.24 + 0.04 0.24 +

0.04

Effective accuracy - Coil only (hall effect clamps)

0.45 + 0.42 0.45 + 0.42 0.45 + 0.42 0.45 +

0.42

Total uncertainty with 3050 (All clamps)

0.48 + 0.44 0.48 + 0.44 0.49 + 0.44 0.53 +

0.45

Total uncertainty with 3041 (wound clamps)

0.25 + 0.06 0.26 + 0.06 0.26 + 0.06 0.28 +

0.07

Total uncertainty with 3041 (hall effect clamps)

0.46 + 0.44 0.46 + 0.44 0.46 + 0.44 0.47 +

0.45

Total uncertainty with 3010 calibrator (wound clamps)

0.25 + 0.042 0.25 + 0.042 0.25 + 0.042 0.26 + 0.042

Total uncertainty with 3010 (hall effect clamps)

0.45 + 0.42 0.46 + 0.42 0.46 + 0.42 0.46 +

0.42

Accuracy is dependant on proper alignment of the clamp meter within the coil

Certain clamp meters have alignment marks which should be aligned with the centre of the coil.

Certain types of clamp meter may have additional errors, or be outside the range which can be driven by the 3041/3010 directly

Uncertainty calculated as the square root of the square of coil ac square of calibrator accuracy 

using empirical data obtained for both wound & hall effect instruments from a wide range of manufacturers

Clamp coil adaptor is supplied complete with workstation incorporating alignment marks (size 275x295x45mm)

EA002 2 / 10 / 50 Turn Clamp Coil Adapter

V4.00

Содержание EA002

Страница 1: ...EA002 2 10 50 TURN 1500A CLAMP COIL ADAPTER OPERATION MANUAL...

Страница 2: ...Version 1 00 April 2007 All product names are trademarks of their respective companies EA002 2 10 50 Clamp Coil Adapter Operation Manual...

Страница 3: ...by the purchaser Transmille will contact the purchaser before proceeding with any repair To obtain repair under this guarantee the purchaser must send the instrument in its original packaging carriag...

Страница 4: ...Terminals to 10 Turn Coil Up to 20A simulation using coil 7 3000 Series 30A Terminals to 10 Turn Coil Up to 300A simulation using coil 7 3000 Series 2A Terminals to 50 Turn Coil Up to 100A simulation...

Страница 5: ...by the number of turns of the coil e g 2 10 and 50 with the Transmille coil The principle is each turn of the coil produces a magnetic field proportional to the current flowing in it If you take 50 w...

Страница 6: ...total accuracy which should be taken into account when calibrating clamp meters using a coil The first is the accuracy of the current produced by the calibrator the second is the coupling between the...

Страница 7: ...MANUAL Transmille Ltd Page 6 2 Turn Coil Minimum 10mm inside jaw diameter 10 Turn Coil Minimum 10mm inside jaw diameter 50 Turn Coil Minimum 25 4mm inside jaw diameter Connection Diagrams Terminal Co...

Страница 8: ...Page 7 3000 Series 2A Terminals to 10 Turn Coil Up to 20A simulation using coil 3000 Series 30A Terminals to 10 Turn Coil Up to 300A simulation using coil 20A output 200A Simulation for 3050 3000 Seri...

Страница 9: ...s 30A Terminals to 50 Turn Coil Up to 1500A simulation using coil 20A output 1000A Simulation for 3050 3000 Series 2A Terminals to 2 Turn Coil Up to 4A simulation using coil 3000 Series 30A Terminals...

Страница 10: ...s all that is recommended CAUTION Do not use cleaning fluids or solvents for cleaning as these may damage the enclosure and affect the plastic materials used in the adapter Handling Precautions The ad...

Страница 11: ...EA002 2 10 50 CLAMP COIL ADAPTER OPERATION MANUAL Transmille Ltd Page 10 EA002 2 10 50 Clamp Coil Adapter Appendix A Specifications...

Страница 12: ...0 41 0 01 0 41 0 01 Effective accuracy Coil only hall effect clamps 0 59 0 11 0 59 0 11 0 59 0 11 0 59 0 11 Total uncertainty with 3050 All clamps 0 61 0 13 0 62 0 13 0 62 0 13 0 65 0 14 Total uncerta...

Страница 13: ...il Only 120uH Coil with typical clamp meter on 50 Turn coil 200uH Coil with typical clamp meter on 10 Turn coil 50uH Coil with typical clamp meter on 2 Turn coil 5uH Specifications apply between 17 C...