EA002 2 / 10 / 50 CLAMP COIL ADAPTER
OPERATION MANUAL
Transmille Ltd.
Page 10
EA002
2 / 10 / 50 Clamp Coil Adapter
Appendix A
Specifications
Страница 1: ...EA002 2 10 50 TURN 1500A CLAMP COIL ADAPTER OPERATION MANUAL...
Страница 2: ...Version 1 00 April 2007 All product names are trademarks of their respective companies EA002 2 10 50 Clamp Coil Adapter Operation Manual...
Страница 3: ...by the purchaser Transmille will contact the purchaser before proceeding with any repair To obtain repair under this guarantee the purchaser must send the instrument in its original packaging carriag...
Страница 4: ...Terminals to 10 Turn Coil Up to 20A simulation using coil 7 3000 Series 30A Terminals to 10 Turn Coil Up to 300A simulation using coil 7 3000 Series 2A Terminals to 50 Turn Coil Up to 100A simulation...
Страница 5: ...by the number of turns of the coil e g 2 10 and 50 with the Transmille coil The principle is each turn of the coil produces a magnetic field proportional to the current flowing in it If you take 50 w...
Страница 6: ...total accuracy which should be taken into account when calibrating clamp meters using a coil The first is the accuracy of the current produced by the calibrator the second is the coupling between the...
Страница 7: ...MANUAL Transmille Ltd Page 6 2 Turn Coil Minimum 10mm inside jaw diameter 10 Turn Coil Minimum 10mm inside jaw diameter 50 Turn Coil Minimum 25 4mm inside jaw diameter Connection Diagrams Terminal Co...
Страница 8: ...Page 7 3000 Series 2A Terminals to 10 Turn Coil Up to 20A simulation using coil 3000 Series 30A Terminals to 10 Turn Coil Up to 300A simulation using coil 20A output 200A Simulation for 3050 3000 Seri...
Страница 9: ...s 30A Terminals to 50 Turn Coil Up to 1500A simulation using coil 20A output 1000A Simulation for 3050 3000 Series 2A Terminals to 2 Turn Coil Up to 4A simulation using coil 3000 Series 30A Terminals...
Страница 10: ...s all that is recommended CAUTION Do not use cleaning fluids or solvents for cleaning as these may damage the enclosure and affect the plastic materials used in the adapter Handling Precautions The ad...
Страница 11: ...EA002 2 10 50 CLAMP COIL ADAPTER OPERATION MANUAL Transmille Ltd Page 10 EA002 2 10 50 Clamp Coil Adapter Appendix A Specifications...
Страница 12: ...0 41 0 01 0 41 0 01 Effective accuracy Coil only hall effect clamps 0 59 0 11 0 59 0 11 0 59 0 11 0 59 0 11 Total uncertainty with 3050 All clamps 0 61 0 13 0 62 0 13 0 62 0 13 0 65 0 14 Total uncerta...
Страница 13: ...il Only 120uH Coil with typical clamp meter on 50 Turn coil 200uH Coil with typical clamp meter on 10 Turn coil 50uH Coil with typical clamp meter on 2 Turn coil 5uH Specifications apply between 17 C...