Faults
Causes
Solutions
Instrument does not start properly
No electricity of the
battery
Charge the battery and observe the
light of the on/off key; if the light is
red, continue to charge; otherwise,
contact the supplier.
Instrument is not charging properly
1.
The
application
environment does not
meet
the
charging
conditions.
2. The battery connection
is bad.
3. Problems of battery or
internal circuit.
1. Charge the instrument at an
environment of 0
℃~
50
℃
.
2. Open the battery compartment
cover to check the battery contacts
and reinstall the battery.
3. Otherwise, contact the supplier
to replace the battery.
The normal curves cannot be tested
1.
The
instrument
parameters are not set
correctly.
2. The optical fiber output
end
face
is
contaminated.
3. The instrument light
output
connector
is
damaged.
4. The
light output
connectors do not match.
1. Set the correct test parameters
again.
2. Clean the light output face.
3. Replace the output connector.
4. Replace the connector with a
matching one.
The test curve has great burrs and
the waveform is not smooth
1. The output connector
is incorrect.
2. The pulse width is set
small.
3. The channels are not
set correctly.
1. Connect appropriate light output
interface again.
2. Increase the value of the test
pulse width.
3. Modify the parameters of the test
channel.
Saturation (flat top) phenomenon
occur in the front of the test curve
1. The pulse width is set
too large.
2. The channel is not set
correctly.
1. Reduce the test pulse width.
2. Modify the parameters of the test
channel.