NOTE 1: TI-500 RFTM will shut down completely within two hours of a lost connection with the 
indicator. 
NOTE: 2: TI-500 RFTM will only enter sleep mode when the TI-700 indicator is OFF. 
 
When shipped from the factory, the TI-700 digital indicator is configured to automatically switch 
off after 5 minutes if not in use. 
 
 

Replacing the batteries – TI-500 RFTM 

 
An optional cylindrical battery holder is available which requires 4 “C” cell alkaline batteries.  
 

1.  Turn the TI-500 RFTM OFF (Press the RED button) 
2.  Locate the metal battery cylinder 
3.  Unscrew either end cap to reveal batteries 
4.  Exchange the batteries – be careful to note polarity 
5.  Replace the end cap 
6.  Re-install the metal battery cylinder 

 
 

Optional Rechargeable battery information 

 

The TI-700 may contain an optional rechargeable battery which is internal to the unit. Before us-
ing the indicator for the first time, please charge the battery overnight. 

The indicator’s battery should operate for about 85 hours if left on continuously. The display unit 
is configured to power down after 5 minutes of inactivity but this parameter can be changed thru 
the A1-10 menu setting. 

The battery can be charged while ON or OFF and the indicator can be operated while it’s charg-
ing unless the state of charge is very low. 

 

WHEN TO CHARGE THE INTERNAL BATTERY 

 

The best time to charge the Lithium Ion battery is any time the indicator is not in use. You need 
not wait for the Low Battery Indication – in fact it’s best that you don’t. Charging the battery 
when not in use keeps the battery “fresh” and is the recommended way to manage it. 

 

When the battery needs to be charged, the Low Battery Indicator will start blinking in the lower 
left-hand corner of the display. The indicator may be used for an additional 2 minutes before it 
automatically powers down.

 It is imperative that you charge the battery at this time to avoid 

permanent damage.

 

 

 

 

Содержание TI-700

Страница 1: ...MODEL TI 700 Digital Weight Indicator with wireless weighing capability User Manual Revision 3 0 July 31 2017...

Страница 2: ...the batteries TI 500 RFTM 5 Optional Rechargeable battery information 5 Balance of this page intentionally left blank 6 DISPLAY KEYPAD DETAILS 7 ADVANCED OPERATION 9 Hold operation 9 NET Weighing 9 Pi...

Страница 3: ...UIPMENT CONTAINS NO USER SERVICEABLE COMPONENTS Servicing of the equipment must only be carried out by trained and authorized personnel Use only the AC adapter supplied with the scale Other adapters m...

Страница 4: ...old with an external TI 500 RFTM Radio Frequency Transceiver Module and an op tional wireless radio your TI 700 indicator is transformed from wired to cable free This config uration is depicted in the...

Страница 5: ...sure press the UNITS key repeatedly un til it is indicated e g lb for pounds kg for kilograms etc The indicator features an automatic zero correction meaning that small deviations will be zeroed autom...

Страница 6: ...he indicator for the first time please charge the battery overnight The indicator s battery should operate for about 85 hours if left on continuously The display unit is configured to power down after...

Страница 7: ...th a pre installed power connector and an external battery charger Simply link the two up and plug the battery charger into a suitable AC wall outlet Note that the external battery charger is rated fo...

Страница 8: ...P Indicates that the indicator is in PEAK HOLD mode lb Indicates that the unit of the displayed weight is in pounds kg Indicates that the unit of the displayed weight is in kilograms g Indicates that...

Страница 9: ...Tare weight has been established first Tare This key is used to establish a Tare weight provided the following conditions are met 1 The indicator is displaying a Gross weight above zero 2 The scale i...

Страница 10: ...values only PEAK HOLD F35 3 The display updates as the load increases but not as the load de creases The value shown on the screen is the maximum weight applied to the scale Press the UNITS key to tog...

Страница 11: ...rs Available choices are 5 10 20 50 and 100 If you continue to push the UNITS key the indicator will resort back to weighing mode and you must start again from Step 2 3 Place the sample items on the p...

Страница 12: ...e The system needs to be returned to the net zero range before another weight can be added 6 The subtotal can be checked by pressing the ON PRINT key for 3 seconds The display shows the net total weig...

Страница 13: ...digit by pressing the UNITS key Decrease the flashing digit by pressing the ZERO OFF key Pressing the TARE key or the ON PRINT key will change the position of the flashing digit 10 After entering the...

Страница 14: ...ht value with one digit blinking The pointer for SP1 is lit 2 Use the four directional keys to adjust the displayed value to the desired low weight value Increase the flashing digit by pressing the UN...

Страница 15: ...SP1 light is ON Backlight colors is BLUE If the displayed weight net or gross is more than the High Max value then Digital Output 3 is activated SP3 light is ON Backlight colors is RED If the displaye...

Страница 16: ...ndicator WIRELESS SYSTEMS ONLY TI 500 RFTM 1 batteries have been depleted and need to be replaced WIRELESS SYSTEMS ONLY TI 500 RFTM 2 batteries have been depleted and need to be replaced WIRELESS SYST...

Страница 17: ...ight Operating temperature 4 F to 104 F 20 C to 40 C Internal connection port for optional printer Tilt adjustable bracket included AC Power 90 308 VAC 50 60 Hz DC Power 5 15 VDC OPTIONAL Bluetooth Sp...

Страница 18: ...d cell damaged on platform Internal fault with indicator call Transcell Tech Support Scale turns off on its own The indicator has a power conservation feature set to automatically power off the scale...

Отзывы: