ACCESSING THE MENUS 

 
To access the Setup (“F) menu: 

1.  Power off the indicator. 

2.  Locate the slide switch on the rear cover and move it to the opposite position. 

NOTE: A metal plate held on by two drilled-head screws may conceal the slide switch. 

3.  Power on the indicator.  

The display shows ” F   1” to indicate that you are in Setup Menu mode. 

4.  Use the navigation keys shown in the figure below to move through the menu. 

 

To access the User (“A) menu: 

1.  Enter the Setup (“F”) menu. 

2.  Use the right or left directional keys shown below to move right or left in the Setup (“F”) menu 

until the indicator shows ” A   1”. 

SETUP MODE KEY FUNCTIONS

ZERO

GROSS

NET

TARE

PRINT

SET

UNITS

 

 

 
MENU STRUCTURE 
 

All menus consist of a top level (heading) and a secondary level. The top level contains the code (e.g. 
F1) for the parameter to be configured. The secondary level contains the selection list or allows access 
to a programming sequence.  
 
Use the directional keys to move around in the Menu Structure shown below. 
 

 

F1 
Grads

 

10

 

80

 

F2 

Sampling

 

Numeric Entry

 

 

Etc. 

 

 

1. 

To move to a new heading, use the TARE (left) or PRINT (right) key to move right or 
left in the Menu. 

 
2. 

To move to the selection or programming level, press the ZERO (down) key once. 
The currently saved selection is shown. 

 
3. 

To view the available selections for the current heading, use the TARE (left) or 
PRINT (right) key to move through the selection field. 

 
4. 

To save a new selection, press the NET/GROSS (Set) key .To exit without saving, 
press the UNITS (up) key to return to the current heading.  

 
5. 

Repeat Steps 2 through 5 until the Menu is programmed.  

 

Содержание TI-500E SS

Страница 1: ...i TI 500E SS Digital Weight Indicator Setup Operation Manual Revision 1 1 July 9 2012...

Страница 2: ...ATION F17 11 VIEW CALIBRATION VALUES F18 12 KEY IN ZERO CALIBRATION VALUE F19 13 KEY IN SPAN CALIBRATION VALUE F20 13 OPERATION 14 DISPLAY 14 KEYBOARD 15 FUNCTION KEYS 15 GENERAL SCALE OPERATION 15 WE...

Страница 3: ...stallation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications...

Страница 4: ...ossible In extreme cases it may be necessary to install surge suppressors line conditioners or even UPS Uninterruptible Power Supplies systems not included Environment Avoid installing the indicator i...

Страница 5: ...on the main PCBA Do not connect the load cell cable shield to the green load cell input terminal The TI 500E SS indicator contains a connection terminal on the main board assembly for connection to t...

Страница 6: ...cord attached to the indicator The internal power supply is rated for 100 240 VAC and 50 60 Hz Simply plug the unit into a standard wall outlet CONFIGURATION OVERVIEW The indicator contains two main...

Страница 7: ...onsist of a top level heading and a secondary level The top level contains the code e g F1 for the parameter to be configured The secondary level contains the selection list or allows access to a prog...

Страница 8: ...ty within which the scale may be zeroed Note that the indicator must be in standstill to zero the scale 100 NA 1 9 2 20 F5 Motion Band Sets the level at which motion is detected If motion is not detec...

Страница 9: ...n value in case of memory loss in the field Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F21 Factory Reset North America This sub menu will reset al...

Страница 10: ...ne 4 20 mA output F23 1 While in the Setup Menu mode scroll to F 23 then scroll down once using the ZERO key to enter Fine tune menu The indicator outputs 4 mA and displays a number 2 While monitoring...

Страница 11: ...t segments decimal points and LCD annunciators in a test sequence Press ZERO key to begin sequence A6 Output String Selects fixed output string for serial port Standard Chipset only Refer to Appendix...

Страница 12: ...he display will show SET momentarily and then revert back up to A8 LF Line Feeds Number Entry A9 1 While in the User Menu mode scroll to A 9 and then scroll down once using the ZERO key to enter the L...

Страница 13: ...troubleshoot setup problems 2 After making sure that there are no test weights on the platform press the ZERO key again to zero out the displayed value 3 Press the NET GROSS key to save the zero point...

Страница 14: ...causes include incorrect load cell wiring a mechanical obstruction or a faulty load cell VIEW CALIBRATION VALUES F18 Note The values displayed in this procedure are valid only after a successful cali...

Страница 15: ...id span calibration values obtained from a successful F17 calibration procedure must be used 1 While in the Setup mode scroll to F 20 and then scroll down once using the ZERO key The indicator will pr...

Страница 16: ...ight is active whenever the displayed weight is within a pre programmed band from true zero NET Indicates that the indicator is displaying net weight GROSS Indicates that the indicator is displaying g...

Страница 17: ...in motion 3 The scale is not in overload see Appendix D for error codes Print This key is used to send weight information out to the serial port provided the following conditions are met 1 The scale...

Страница 18: ...u continue to push the UNITS key the indicator will resort back to weighing mode and you must start again from Step 2 3 Place the sample items on the platform all at once and allow the weight indicati...

Страница 19: ...usly and at the same transmission baud rate configured in A1 LEGAL FOR TRADE SEALING Indicators can be sealed for commercial Legal for Trade applications as follows 1 Power off the indicator 2 On the...

Страница 20: ...hipset 1024 x 8 internal to C Enhanced Chipset EEPROM 256 x 8 external to C Digital Filtering Software selectable SERIAL COMMUNICATIONS Serial Port Full Duplex selectable Baud rate 8 data bits no pari...

Страница 21: ...tor to zero the scale The indicator will not respond if the scale is in motion positive overload or negative overload The indicator will also not respond if it is not in gross mode or within the zero...

Страница 22: ...ondec Demand String String Format 1 is designed for two way communication STX POL xxxxx xx LB KG GR NT CR LF Polarity SP Positive Negative Weight Data Units LB pound KG kilogram Gross Net GR Gross NT...

Страница 23: ...Keyed in weight value is less than 1 of full scale capacity Use a larger test weight or check keyed in value Err 2 Span Calibration Mode F17 There is not enough load cell signal to produce the intern...

Отзывы: