14
CODE/NAME
DESCRIPTION
SELECTION LIST
F31
Gross Zero
Band
Selects the range within which the scale will automatically clear the
tare and switch to Gross mode. Note that the scale must be in stand-
still. Selections are in display divisions (d). Scrolling down with the
ZERO key one level begins the procedure.
“0” = Disabled
Key-in
0 - 10
0
√√√√
F32
Center of Zero
Band
Selects the range around gross zero within which the scale will dis-
play the Center of Zero annunciator. Selections are in display divi-
sions (d).
0.25d
√√√√
0.5
F34
Auto Print Min.
Weight
Selects the minimum weight at which the auto print function will work
if enabled. Selections are in display divisions (d). Scrolling down with
the ZERO key one level begins the procedure.
“0” = Disabled
Key-in
0 - 100
1
√√√√
Fine-tune 4-20 mA output (F24)
1. While in the Setup Menu mode, scroll to "F 24", then scroll down once using the
ZERO key to enter Fine-tune menu. The indicator outputs 4 mA and displays a num-
ber.
2. While monitoring the voltage across R
L
(see Appendix), use the right (PRINT) or left
(TARE) keys to change the displayed value until the measured voltage is exactly 1
VDC.
3. Press the SET (Net/Gross) key to save. The indicator outputs 20 mA and displays an-
other number.
4. While monitoring the voltage across R
L
, use the right (PRINT) or left (TARE) keys to
change the displayed value until the measured voltage is exactly 5 VDC.
5. Press the SET (Net/Gross) key to save and revert back to F24.