4

Calibration

All SRP Series scales may be calibrated with any precision test weight from 10% to 100% of the

scale’s capacity. The recommended test weight is about 2/3 of the full-scale capacity, but in
general, the larger the test weight, the more accurate the scale will be. The only restriction is that
the test weight 

must

 be in pounds.

1. Turn the scale off. Locate the hidden switch on the base panel and set it at the rightmost po-

sition. Turn the unit back on. The message "

C 0

" appears on the display briefly, followed by a

value which remains on the screen. Allow a 20 minute warm-up period for the load cell and
indicator to become thermally stable.

2. Press 

ZERO

 to zero the value, then press the 

NET/GROSS

 key to save the zero point value.

3. The display will momentarily prompt "

C 1

" for the span calibration, followed by "

0. 0

" with one

digit flashing. Place the test weight on the platform.

4. Use the four directional keys to adjust the displayed value to the actual test weight value in

pounds. Increase the flashing digit by pressing the 

lb/kg

 key. Decrease the flashing digit by

pressing the 

ZERO

 key. Pressing the PRINT key or the TARE key can change the position of

the flashing digit.

5. After setting the exact value, press the 

NET/GROSS

 key to save the value.

6. If the calibration was successful, the display will show "

ECAL

" momentarily, then freeze. Exit

the Calibration mode and enter the Normal Operating Mode by turning off the scale, posi-
tioning the calibration switch back to the leftmost position, and turning the scale back on.

7. If the calibration was 

not

 successful, one of the error messages below will appear. Take the

indicated action to correct the problem, then perform a new calibration.

"Err0"

 - The calibration test weight or the adjusted keyed-in weight is larger than full scale.

Change the calibration test weight or check the keyed-in weight.

"Err1"

 - The calibration test weight or the adjusted keyed-in weight is smaller than 10% of full

scale. Change the calibration test weight or check the keyed-in weight.

"Err2"

 - Check keyed-in weight with the actual weight placed on platform.

Displayed Error Messages

Display Message

Cause or Condition

-HI-

Scale was turned on with a weight on the platform that exceeded the Zero
Capture Range. Scale must be turned on without extra weight on the plat-
form.

-LO-

Scale was turned on without stainless steel platform on scale base that ex-
ceeded the Zero Capture Range in the negative direction. Replace platform
and restart.

r

 

r

 

r

 

r

 

r

Total weight on platform exceeds scale's capacity.

– – – – –

During normal operation, the stainless steel platform was removed, resulting
in an under range condition. Replace scale platform.

Содержание SRP-300

Страница 1: ...SRP 300 Digital Bench Scale Operation Manual Revision 1 2 August 24 2000...

Страница 2: ...ell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60090 Tel 847 419 9180 Fax 847 419 1515 E mail transcell transcell net Web...

Страница 3: ...communication parameters that are embed ded in the User Menu The User Menu is accessed through the front panel keys Complete directions in cluding a graphical road map are found in the Configuration s...

Страница 4: ...separate menu selections each with its own sub menu of choices If your scale is con nected by RS 232 serial communication lines to other equipment or you wish to dis able the lb kg key you may have t...

Страница 5: ...t and one stop bit 7O 7 data bits with odd parity bit and one stop bit 7E 7 data bits with even parity bit and one stop bit 7 7 data bits with no parity bit and two stop bits 8 7O 7E 7 A3 Mode of Seri...

Страница 6: ...e NET GROSS key to save the value 6 If the calibration was successful the display will show ECAL momentarily then freeze Exit the Calibration mode and enter the Normal Operating Mode by turning off th...

Страница 7: ...0 C to 40 C OPTIONAL SWIVEL POST 18 Tall Stainless Steel Construction POWER SOURCE AC Adapter 12VDC 800mA included WEIGHT Net Weight 45 lb 20 4 kg Shipping Weight 50 lb 22 7 kg Certificate of Conforma...

Страница 8: ...the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper or unau thorized repair or improper testing and 4 Such goods have not been mod...

Отзывы: