Table of Contents

Introduction ........................................................................................ 1
Installation and Wiring........................................................................ 1
Keyboard Functions ........................................................................... 2
Configuration...................................................................................... 2
User Menu Descriptions ..................................................................... 3
Calibration .......................................................................................... 4
Displayed Error Messages ................................................................. 4
Specifications ..................................................................................... 5
Warranty and Service Information...................................................... 6

Introduction

The SRP Series Digital Bench Scale is a
large (18” x 18”) bench scale with remote
digital indicator. The system is NTEP-
certified for Class III at 3,000 divisions, and
is legal-for-trade. Housed in a standard
black polymer case, the indicator displays
its numerals on a bright VFD screen. The
base unit features a removable stainless
steel platform, an RS-232 serial communi-
cation port, and adjustable feet for leveling.

The SRP Series uses full duplex RS-232
serial format for communication with many
types of attached support equipment. The

unit can transmit data on demand, or con-
tinuously in several popular data protocols
to match a wide variety of printers, remote
displays, or personal computers. If you plan
to use your scale with equipment other than
a PC, you may have to alter the serial
communication parameters that are embed-
ded in the User Menu.
The User Menu is accessed through the
front panel keys. Complete directions, in-
cluding a graphical road map, are found in
the Configuration section of the manual.

Installation and Wiring

Unless you wish to interface to a device
other than a PC, no wiring is necessary.
Simply connect the indicator to the scale
platform, plug in the AC adapter, and turn
the unit on. If the swivel indicator option was
ordered, the stainless steel post may be
mounted to the bottom of the base with the
hardware provided. Alternately, it may be
mounted to a table or bench. In general,

please allow a 20-minute warm up period
before using the scale.

If interfacing to a PC, simply connect the
optional null modem cable to a serial port
(e.g. COM 1) on your PC. If your computer
has a 25-pin serial port, a standard 9-pin to
25-pin adapter, such as the type used for a
mouse, may be used and is readily avail-
able at your local computer supply retailer.

If interfacing to a serial device other than a PC, you may have to
wire your own cable to communicate with that device. If so, refer
to the 9-pin connector pin out shown at right.

Pin No.

Pin Name

2

Receive Data

3

Transmit Data

5

Signal Ground

Содержание SRP-300

Страница 1: ...SRP 300 Digital Bench Scale Operation Manual Revision 1 2 August 24 2000...

Страница 2: ...ell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60090 Tel 847 419 9180 Fax 847 419 1515 E mail transcell transcell net Web...

Страница 3: ...communication parameters that are embed ded in the User Menu The User Menu is accessed through the front panel keys Complete directions in cluding a graphical road map are found in the Configuration s...

Страница 4: ...separate menu selections each with its own sub menu of choices If your scale is con nected by RS 232 serial communication lines to other equipment or you wish to dis able the lb kg key you may have t...

Страница 5: ...t and one stop bit 7O 7 data bits with odd parity bit and one stop bit 7E 7 data bits with even parity bit and one stop bit 7 7 data bits with no parity bit and two stop bits 8 7O 7E 7 A3 Mode of Seri...

Страница 6: ...e NET GROSS key to save the value 6 If the calibration was successful the display will show ECAL momentarily then freeze Exit the Calibration mode and enter the Normal Operating Mode by turning off th...

Страница 7: ...0 C to 40 C OPTIONAL SWIVEL POST 18 Tall Stainless Steel Construction POWER SOURCE AC Adapter 12VDC 800mA included WEIGHT Net Weight 45 lb 20 4 kg Shipping Weight 50 lb 22 7 kg Certificate of Conforma...

Страница 8: ...the nonconformity actually exists and was not caused by accident misuse neglect alteration improper installation improper or unau thorized repair or improper testing and 4 Such goods have not been mod...

Отзывы: