Factory default control parameters
The parameters used in the controller have been developed by Accurate Thermal Systems to give the best unit
performance for most applications. If the need arises the “AT” autotune command can be run to further optimize
results based on the thermal block load and ambient conditions. Contact us for further support and details.
Switching the display from Degrees C to Degrees F and vice versa
To switch the display press and hold down the left key until the display indicates “CN-t”, next press the 2
nd
key
from the left one time and the top line will show “d-U” which is display units. Switch to either C or F. Based on
which setting you use the parameters below must be changed to those shown. They will display by pressing the
2
nd
key from the left after setting C or F. After all values have been changed hold down the left most key so the
controller resets and accepts the new values.
Parameter
Degrees C
Degrees F
SL-H
410.0
770.0
SL-L
0.0
0.0
To correct the calibration offset – press button on left, then button 2
nd
from the left until the parameter CN5 is displayed
CN5
If in C multiply value
by 1.8 to switch to F
If in F divide value
by 1.8 to switch to C