Transcat ThermCal400 Скачать руководство пользователя страница 2

The ThermCal400 

 
Introduction 

 
The ThermCal400 calibrator provides a safe, dry, constant temperature source for checking and calibrating a wide 
range of temperature sensors, systems, indicators and thermometers It is fast and economical and can be used 
either on a bench top or as a portable field unit. The weight of the unit is only 11 pounds/five  kilograms.  The 
unit covers the temperature range from 5°C above ambient up to 400°C using a machined aluminum block as the 
heat transfer medium. The temperature control circuit is built into the unit. 
 
 

Features include: 

• Maximum temperature of 400°C/752°F 
• An independent over-temperature cutout 
• Up to eight setpoints can be stored & recalled 

 
 
Even though the unit heats up rapidly, highly efficient insulation and an internal cooling fan ensures that the case 
remains cool enough to handle even at maximum operating temperatures.  The ThermCal400 calibrator has been 
designed to comply with all relevant electromagnetic interference and electrical safety regulations. 

 
 
 
 
 

-1- 

Содержание ThermCal400

Страница 1: ...ThermCal400 Dry Block Temperature Calibrator INSTRUCTION MANUAL Please read all the information in this booklet before using the unit August 2013...

Страница 2: ...above ambient up to 400 C using a machined aluminum block as the heat transfer medium The temperature control circuit is built into the unit Features include Maximum temperature of 400 C 752 F An inde...

Страница 3: ...y 0 015 C at 200 C 0 025 C at 400 C Heat up time 20 C to 300 C 9 minutes Cool down 300 C to 100 C 18 minutes Immersion Depth 4 5 114 3mm Fan Cooling Automatic Weight 11 lbs 5 Kg Dimensions H x W x D 8...

Страница 4: ...perate the unit close to inflammable liquids or gases DO NOT place any liquid directly in your unit At all times USE COMMON SENSE Operator Safety All operators of Accurate Thermal Systems equipment mu...

Страница 5: ...with the following value fuse to protect the cable 230V UK 4 AMP The fuse in the unit protects the unit and the operator Note that units marked 230V on the rating plate work at 220V units marked 120V...

Страница 6: ...etained 2 When you have the correct set temperature displayed the unit will start to heat or cool to that value 3 Once the process value actual temperature reaches the set point allow the block to ful...

Страница 7: ...ay indicates CN t next press the 2nd key from the left one time and the top line will show d U which is display units Switch to either C or F Based on which setting you use the parameters below must b...

Страница 8: ...packed for shipment If in doubt contact your supplier 1 Cleaning Before cleaning your unit ALWAYS disconnect it from the power supply and allow it to cool below 50 C Your unit can be cleaned by wipin...

Страница 9: ...carrying strap 4153 Insert extractor ATS3041 insert 1 8 3 16 5 16 3 8 ATS3047 Blank insert ATS3043 Insert 5 x 1 4 ATS3048 insert 1 x 9 16 1 x 1 4 ATS3044 Insert 2 x 1 4 2 x 3 8 ATS3049 insert 1 x 5 8...

Страница 10: ...isuse accident or negligence Servicing under this guarantee should be obtained from the supplier Not withstanding the description and specification s of the units contained in the Operator s Manual Ac...

Отзывы: