Replacement Parts 

The following parts may be obtained from Anville Instruments Ltd. if replacements or alternatives are required: 

Part Number  

 

Description

 

 4163    

 

UK 240 volt power cable with 13amp UK plug (5 amp fuse) 

 4164 

 

 

Euro style 240 volt power cable with R/A Schuko plug 

 4150 

 

 

US style 120 volt power cable 

 4282 

 

 

Instruction manual 

 4150 

 

 

Unit carrying strap 

 4285 

 

 

Insert extractor 

 ATS3074 

 

insert for 1/8” diameter probes 

 

 ATS3075 

 

Insert for 3/16” diameter probes 

 

 ATS3076        

 

Insert for ¼” diameter probes 

 

 ATS3077 

 

Insert for 5/16” diameter probes 

 

 ATS3078 

 

Insert for 3/8” diameter probes 

 

  

Spare Parts 

Part Number 

 

Description 

 

  4224   

 

Peltier 

  4174   

 

Temperature controller 

  4147    

 

PRT 

  4221   

 

Solid state relay 

  4223   

 

Power supply 

  4280   

 

3 PDT power relay 

  

4283 

 

 

2 amp fuse, 5 x 20mm

 

 

-8- 

Содержание ThermCal130

Страница 1: ...ThermCal130 Dry Block Temperature Calibrator INSTRUCTION MANUAL Please read all the information in this booklet before using the unit January 2015 Rev 1 0 ...

Страница 2: ...of 20 C using a machined aluminum block as the heat transfer medium The temperature control circuit is built into the unit Features include Maximum temperature of 130 C 266 F Minimum temperature of 20 C 4 F Up to eight setpoints can be stored recalled insert well diameter reference well Even though the unit heats up and cools down rapidly highly efficient insulation and an internal cooling fan ens...

Страница 3: ...tes Cool down 20 to 20 C 9 minutes ambient of 20 C Cool down 130 to 20 C 4 minutes Immersion Depth 4 101mm Insert well diameter x 4 depth Reference well diameter x 4 depth Fan Cooling Automatic Weight 17 lbs 7 7 Kg Dimensions H x W x D 11 x 9 x 8 inches 279 x 229 x 203 mm excluding the carrying strap Electrical supply Voltage Cycles Power 230V 50 60Hz 150W 120V 50 60Hz 150W Note The above specific...

Страница 4: ...lose to inflammable liquids or gases DO NOT place any liquid directly in your unit At all times USE COMMON SENSE Operator Safety All operators of Accurate Thermal Systems equipment must have available the relevant literature needed to ensure their safety It is important that only suitably trained personnel operate this equipment in accordance with the instructions contained in this manual and with...

Страница 5: ...d with the following value fuse to protect the cable 230V UK 4 AMP The fuse in the unit protects the unit and the operator Note that units marked 230V on the rating plate work at 220V units marked 120V work at 110V In both cases however the heating rate will degrade by approximately 8 The rating plate is on the rear of the unit 3 Plug the power cable into the socket on the rear of the unit 4 Place...

Страница 6: ...t to heat or cool to that value 3 Once the process value actual temperature reaches the set point allow the block to fully stabilize for at least 10 minutes before performing a calibration 4 When calibrating sensors and thermometers a best practice method is to start at higher temperatures first and then work down to lower ones This will allow devices to be removed safely Entering up to 8 setpoint...

Страница 7: ...tune command can be run to further optimize results based on the thermal block load and ambient conditions Contact us for further support and details Switching the display from Degrees C to Degrees F and vice versa To switch the display press and hold down the left key until the display indicates CN t next press the 2nd key from the left one time and the top line will show d U which is display uni...

Страница 8: ...rly packed for shipment If in doubt contact your supplier 1 Cleaning Before cleaning your unit ALWAYS disconnect it from the power supply and allow it to cool to 30 C Your unit can be cleaned by wiping with a damp soapy cloth Care should be exercised to prevent water from running inside the unit Do not use abrasive cleaners 2 Fuses Your unit is protected by fuse They should only be changed by suit...

Страница 9: ...50 US style 120 volt power cable 4282 Instruction manual 4150 Unit carrying strap 4285 Insert extractor ATS3074 insert for 1 8 diameter probes ATS3075 Insert for 3 16 diameter probes ATS3076 Insert for diameter probes ATS3077 Insert for 5 16 diameter probes ATS3078 Insert for 3 8 diameter probes Spare Parts Part Number Description 4224 Peltier 4174 Temperature controller 4147 PRT 4221 Solid state ...

Страница 10: ...e result of misuse accident or negligence Servicing under this guarantee should be obtained from the supplier Not withstanding the description and specification s of the units contained in the Operator s Manual Accurate Thermal Systems hereby reserves the right to make such changes as it sees fit to the units or to any component of the units This manual has been prepared solely for the convenience...

Отзывы: