For example

:

If the measured environment temperature is 25ºC, rela-
tive humidity is 50%, dew point is 13.8ºC, then:

Green light: when the measured surface temperature is 
between 77ºF (25ºC) - 70.3ºF (21.3ºC);
Yellow light: when the measured surface temperature is 
between 70.2ºF (21.2ºC) - 63.5ºF (17.5ºC);
Red light: when the measured surface temperature is 
between 63.4ºF (17.4ºC) - 56.8ºF (13.8ºC).

NOTE - CONCRETE MOISTURE AND DEW POINT 

It is recommended that the temperature of the surface 
of the concrete should be at least 10ºF (5ºC) above the 
ambient dew point temperature at the time of installation 
so as to avoid condensation related issues.

BATTERY REPLACEMENT

• 

When battery power is not sufficient, LCD will 
display replacement with one new battery type 9V 
is required.

• 

Open battery cover, take the battery out of the 
instrument and replace with a new 9-Volt battery, 
then replace the battery cover.

7

Содержание IRT2DP

Страница 1: ...IRT2DP Dual Laser IR Thermometer and Dew Point Detector USER GUIDE...

Страница 2: ...ront Panel Controls and Indicators Laser Beam Backlight Mode Surface Temperature Dew Point DP Surface Temperature Ambient Temperature AT Surface Temperature Relative Humidity RH Measurement Operation...

Страница 3: ...ty range is from 0 100 RH The dew point temperature range is 30 to 100 C 22 to 212 F see Specifications for details The IRT2DP has a fast response time 0 15 sec and is powered with a 9V battery and co...

Страница 4: ...TING INSTRUCTIONS 1 Switch Button automatically switch off at about 15 minutes 2 Backlight Laser Button 3 Mode Button Surface temperature with RH Dew Point or AmbientTemperature 4 C F Audio On Off But...

Страница 5: ...once to turn ON Press the measurement trigger again to turn OFF SWITCHING C F Select the temperature units C or F pressing once the C F button AUDIO SIGNAL ON OFF Press the C F button for 3 seconds t...

Страница 6: ...1 DP Dew Point 2 AT AmbientTemperature 3 RH Relative Humidity 4 Laser ON symbols 5 Low Battery Indication 6 RH Indicator bar Proximity to dew point indicator bar 7 Emissivity symbol and value 8 C F sy...

Страница 7: ...t ON laser beam OFF backlight ON MODE After switch on the instrument will be in the same mode as the last time used It will detect and display the surface temperature with either dew point temperature...

Страница 8: ...icates that the temperature and humidity are normal Yellow Light If the measured surface temperature is within the 2 3 range of ambient temperature to dew point temperature the yellow light and interm...

Страница 9: ...etween 63 4 F 17 4 C 56 8 F 13 8 C NOTE CONCRETE MOISTURE AND DEW POINT It is recommended that the temperature of the surface of the concrete should be at least 10 F 5 C above the ambient dew point te...

Страница 10: ...or disassemble the instrument or immerse it in water Repairs or service are not covered in this manual and should only be carried out by qualified trained tech nician For service use only manufacturer...

Страница 11: ...5 C 20 to 380 C 32 to 716 F 1 5 C Wavelength range 8 14um Humidity Range 0 100 RH Humidity accuracy 3 5 20 to 80 Dew point temperature 30 to 100 C 22 to 212 F D S 20 1 Operating temp 0 to 50 C 32 to 1...

Отзывы: