INTRODUCTION

The IRT2DP is a handheld dual laser, non-contact 
Infrared Thermometer and Dew Point detector with 
20:1 distance to spot ratio, tricolor light (and audible) 
indication as reading goes close to dew point tempera-
ture. It is used to measure surface temperature along 
with relative humidity, dew point temperature and 
ambient temperature. 
The temperature range is -50°C to 380°C (-58ºF to 
716ºF) and has an accuracy of approximately 1 to 
1.5%. The humidity range is from 0 – 100% RH. The 
dew point temperature range is  -30º to 100ºC (-22º 
to 212ºF) (see Specifications for details). The IRT2DP 
has a fast response time (0.15 sec) and is powered with 
a 9V battery and complies with Class II UK/EU laser 
safety standard EN60285.

KEY FEATURES

•   Rapid detection function
•   Precise non-contact measurements
•   Dual laser targeting
•   3 Mode display - Relative Humidity / Dew Point 
     Temperature / Ambient Temperature
•   Green  - Yellow – Red LED light and audible signal  
     indicate proximity to Dew Point Temperature
•   Automatic Data Hold 
•   User selectable ºC or ºF
•   Pre-set Emissivity of 0.95
•   Backlight LCD display
•   Unique flat surface, modern housing design

1

Содержание IRT2DP

Страница 1: ...IRT2DP Dual Laser IR Thermometer and Dew Point Detector USER GUIDE...

Страница 2: ...ront Panel Controls and Indicators Laser Beam Backlight Mode Surface Temperature Dew Point DP Surface Temperature Ambient Temperature AT Surface Temperature Relative Humidity RH Measurement Operation...

Страница 3: ...ty range is from 0 100 RH The dew point temperature range is 30 to 100 C 22 to 212 F see Specifications for details The IRT2DP has a fast response time 0 15 sec and is powered with a 9V battery and co...

Страница 4: ...TING INSTRUCTIONS 1 Switch Button automatically switch off at about 15 minutes 2 Backlight Laser Button 3 Mode Button Surface temperature with RH Dew Point or AmbientTemperature 4 C F Audio On Off But...

Страница 5: ...once to turn ON Press the measurement trigger again to turn OFF SWITCHING C F Select the temperature units C or F pressing once the C F button AUDIO SIGNAL ON OFF Press the C F button for 3 seconds t...

Страница 6: ...1 DP Dew Point 2 AT AmbientTemperature 3 RH Relative Humidity 4 Laser ON symbols 5 Low Battery Indication 6 RH Indicator bar Proximity to dew point indicator bar 7 Emissivity symbol and value 8 C F sy...

Страница 7: ...t ON laser beam OFF backlight ON MODE After switch on the instrument will be in the same mode as the last time used It will detect and display the surface temperature with either dew point temperature...

Страница 8: ...icates that the temperature and humidity are normal Yellow Light If the measured surface temperature is within the 2 3 range of ambient temperature to dew point temperature the yellow light and interm...

Страница 9: ...etween 63 4 F 17 4 C 56 8 F 13 8 C NOTE CONCRETE MOISTURE AND DEW POINT It is recommended that the temperature of the surface of the concrete should be at least 10 F 5 C above the ambient dew point te...

Страница 10: ...or disassemble the instrument or immerse it in water Repairs or service are not covered in this manual and should only be carried out by qualified trained tech nician For service use only manufacturer...

Страница 11: ...5 C 20 to 380 C 32 to 716 F 1 5 C Wavelength range 8 14um Humidity Range 0 100 RH Humidity accuracy 3 5 20 to 80 Dew point temperature 30 to 100 C 22 to 212 F D S 20 1 Operating temp 0 to 50 C 32 to 1...

Отзывы: