Intended Use

The Trak

®

 Male Fertility Testing System is intended for semi-quantitative 

assessment of sperm concentration at 15 million sperm per milliliter 
(M/mL) or below, between 15 and 55 M/mL, and above 55 M/mL. 
The System also provides a qualitative assessment of semen volume.

Sperm concentration and semen volume are only two factors that 
could impact a man's fertility status and time to pregnancy. For 
complete assessment of male reproductive health, the user should 
consult a physician. For 

in vitro

, over the counter home use.

How it Works

The Trak sperm concentration test uses centrifugal force to isolate 
sperm cells from a semen sample. Any sperm cells present form a white 
column in the Prop channel. The height of the column correlates with 
the concentration of the sperm cells in the sample. The Trak Volume Cup 
accumulates semen into a defined space to indicate semen volume.

Semen Volume

Low (1.5 mL or below)

Normal (Above1.5 mL)

Sperm Concentration

Low (15 M/mL or below)

Moderate (Between 15 & 55 M/mL)

Optimal (Above 55 M/mL)

3

Содержание Trak Male Fertility Testing System

Страница 1: ...OWNER S GUIDE...

Страница 2: ...for Collecting a Sample Running the Test Check for Potential Errors Understand Your Sperm Concentration Result After Testing Cleaning Disinfection Storage Running a Control Test Test Accuracy Frequen...

Страница 3: ...ed to prove paternity This system cannot be used to confirm the success of a vasectomy or as a method of birth control This system does not protect against sexually transmitted diseases DO NOT RE USE...

Страница 4: ...assessment of male reproductive health the user should consult a physician For in vitro over the counter home use How it Works The Trak sperm concentration test uses centrifugal force to isolate sperm...

Страница 5: ...ng a control test prior to running a real test Super Sani Cloth wipes are recommended to disinfect your Engine after use and are available for purchase online Please allow 45 minutes after collecting...

Страница 6: ...ERS BOTTLE OF CONTROL SOLUTION A 4 SAMPLE VOLUME CUPS WITH LIDS If any components of the system are missing or damaged contact customer service Monday Friday 8 00am 6 00pm PST at 1 888 241 2476 2 AA B...

Страница 7: ...Engine Prop Details RED YELLOW LIGHT ENGINE GREEN LIGHT BASE LID INSTALL BATTERIES UNDERNEATH METAL ENGINE ROD HUB S I D E V I E W PROP CENTER HOLE CHANNEL 6...

Страница 8: ...O WITHOUT EJACULATING FOR 3 DAYS BEFORE COLLECTING A SAMPLE THE NEXT TIME AT LESS THAN 2 DAYS SPERM CONCENTRATION MAY BE LOW AT MORE THAN 7 DAYS SPERM CONCENTRATION MAY BE HIGH AND FULL OF DEAD CELLS...

Страница 9: ...Y AND WITHIN REACH MOST SPERM LIVE IN THE 1STDROPS O F E J A C U L A T E NEITHER SHOULD BE USED WHEN COLLECTING A SAMPLE COLLECTION VIA SEXUAL INTERCOURSE IS ALSO NOT RECOMMENDED M A T E R I A L S I N...

Страница 10: ...soap and water before and after handling the sample Volume Cup dropper Prop or Engine Review the guidelines on page 7 for collecting a semen sample to ensure an accurate result If you have never compl...

Страница 11: ...02 Swirl the cup for at least 15 seconds 15 SECS Collect a Sample 10...

Страница 12: ...ook straight at the cup 04 03 You need to wait 30 minutes for the sample to liquefy at room temperature before running the test The test can be performed up to 2 hours after sample collection 30 MINS...

Страница 13: ...y curve up at the edges Record your semen volume result before testing sperm concentration Read and Record Semen Volume Result Men with low semen volume 1 5 mL may be at risk of infertility and or uro...

Страница 14: ...Running the Test After reading the volume result swirl the sample Volume Cup again for at least 15 seconds to evenly distribute the sperm cells Remove the lid from the sample Volume Cup 06 15 SECS 13...

Страница 15: ...move the Prop Remove the foil seal from the Prop Discard pouch and foil seal into the empty test bag 08 Remove the contents from one test bag Prep the Sample An extra dropper and sticker are included...

Страница 16: ...tick part of the dropper is completely full If a whole sample is collected but there isn t enough to fill the dropper discard the cup and dropper and try again in 2 to 7days If your sample is too thic...

Страница 17: ...e open tip of the dropper into the center hole of the Prop Squeeze the top bulb to empty the sample The bottom bulb of the dropper will not empty Use one dropper load only Discard dropper into an empt...

Страница 18: ...Running the Test 12 Place the Engine on a flat stable surface Open the lid Pick up the Prop Attach the Prop to the Engine by pressing it firmly onto the metal Engine rod 17...

Страница 19: ...ight flash The Engine didn t detect the Prop and will not start Try to push the Prop on more securely If the Prop will not attach securely you may need to start over with a new Prop dropper sticker an...

Страница 20: ...red light indicates an error for example the lid was opened before the test was complete and the test result is not valid Repeat the test with a new Prop dropper sticker and any remaining sample in t...

Страница 21: ...s no error see page23 for details on reading your result Read Your Sperm Concentration Result Read Your Test Result Looking at the Prop channel at an angle can cause the white column to appear taller...

Страница 22: ...reason for air bubbles is not adding enough sample to the center hole of the Prop Make sure you add one full dropper load of sample into the Prop EXAMPLE OF AIR BUBBLE Make sure there are no air bubb...

Страница 23: ...on for clogs is not swirling the sample well Make sure you swirl the sample for at least 15 seconds before filling the dropper Make sure there is no red at the bottom of the white column If your Prop...

Страница 24: ...e of the Trak system Refer to page27 If there is no white column the second time you run a test your sperm concentration is low and you are advised to consult a physician Refer to the FAQs on page 39...

Страница 25: ...N MILLIONS PER MILLILITER M mL Moderate Between 15 55 M mL Men with sperm concentrations 55 M mL or below may take longer to conceive A white column that falls on the15 or 55 M mL line should be inter...

Страница 26: ...roppers and stickers can be placed in the empty test bag and Volume Cups can be re capped These items can be put into everyday trash Do not get any liquids dirt dust semen or control solution inside t...

Страница 27: ...two 2 minutes Do not use other disinfecting products Disinfect the Engine if it is exposed to semen Disinfect the Engine before allowing anyone else to handle the Engine When finished thoroughly wash...

Страница 28: ...ror and possibly inaccurate test result Tips or additional information Perform a control test with the provided bottle of control solution to ensure proper operation of the Engine and Props Use the co...

Страница 29: ...02 Open the Prop pouch and remove the Prop Remove the foil seal from the Prop Discard the pouch and foil seal into the empty test bag 28...

Страница 30: ...04 Pick up the dropper Squeeze the top bulb of the dropper and then insert the open tip of the dropper into the deepest part of the control solution bottle Release the top bulb to fill Make sure that...

Страница 31: ...not empty Insert the open tip of the dropper into the center hole of the Prop Squeeze the top bulb to empty the solution Use one dropper load only Cap control solution bottle and place in Trak kit Dis...

Страница 32: ...rol Test 07 Place the Engine on a flat stable surface Open the lid Pick up the Prop Attach the Prop to the Engine by pressing it firmly onto the metal Engine rod 06 Seal the Prop with a seal before sp...

Страница 33: ...teries are low Single red light flash The Engine didn t detect the Prop and will not start Try to push the Prop on more securely If the Prop will not attach securely you may need to start over with a...

Страница 34: ...nger than three minutes the red light will illuminate This means the result is invalid and you will need to run another control test with a new Prop dropper and sticker Blinking red light A blinking r...

Страница 35: ...esult Looking at the Prop channel at an angle can cause the white column to appear taller or shorter than it actually is See page 35 for more details on reading your control test result Look straight...

Страница 36: ...Running a Control Test Expected Control Test Result Locate the top of the white column in the Prop channel WHITE COLUMN 35...

Страница 37: ...like the expected result image a second time do not use the Trak testing system Call customer service at 1 888 241 2476 For Control Solution A the height of the white column should be just above the...

Страница 38: ...93 3 15 55 M mL 82 4 55 M mL 95 5 Factors that may impact test accuracy include Presence of white blood cells most commonly occurring due to infection or recent surgery Abstaining for less than 2 days...

Страница 39: ...uidelines for running the test Truly abstain make sure to collect a full sample and swirl the sample properly in the Volume Cup If you need more help visit trakfertility com or contact customer servic...

Страница 40: ...ups are specially treated with an enzyme Chymotrypsin to reduce the viscosity of the semen sample Most samples will be thin enough to fill the dropper within 30 minutes after collection If your sample...

Страница 41: ...may be misinterpreted if the Prop is viewed at an angle other than straight down because the sperm cells are contained inside the Prop while the15 and 55 orange lines are printed on top of the Prop W...

Страница 42: ...lt above 15 M mL indicate that I am fertile Not necessarily Sperm concentration is an important indicator of male reproductive health but there are other factors including sperm motility sperm morphol...

Страница 43: ...possible risk for infertility Low semen volume and low sperm concentration Low semen volume and moderate sperm concentration Low semen volume and optimal sperm concentration Normal semen volume and lo...

Страница 44: ...Manufactured by Sandstone Diagnostics 6111J Southfront Rd Livermore CA 94551 P A T E N T S P E N D I N G trakfertility com 1 888 241 2476 support trakfertility com TTI 0 2 R3...