8. Connect the pressure hose to the hand pump's MINIMESS
®
connection,
and tighten the connection.
9. Tightly screw the MINIMESS
®
adapter on the pressure hose.
10. Select the suitable adapters and seals for the test item's connection.
11. Screw the adapter for the test sample onto the MINIMESS
®
adapter.
12. Connect the test sample with seal to the adapter.
IMPORTANT! Do not yet tighten the test item!
13. Now pump until the hydraulic liquid escapes at the test sample's
connection. The pressure hose and the test sample connection are now
ventilated.
CAUTION! Demineralised water!
During the quality inspection of the TRAQC-5 P1000, a functional test is
carried out with demineralised water. The reservoir may therefore contain
residues of this liquid.
Check the compatibility with your application and remove the residue
using appropriate measures (e.g. rinsing out with the hydraulic liquid).