Trace2O HydroTest HTCOD-TR Скачать руководство пользователя страница 8

 

 

4.

 

F

u

n

cti
o

n

 schem

at

ics

 

8

 

 

 
 
 

After  switching  on  the  instrument  with  the  main  on/off  switch,  the  heater  is  not  active.  Press  the  HEAT-key  for  heating  up  (status  1). 
After  heating  up  to  the  selected  temperature  the  timer  starts  after  pressing  START-key  (status  3). 

 

Status 

Heat-LED 

Temp.-LED 

Time-LED 

Heat-key 

Start-key 

Temp-/Time- 
key 

Heater 

next  possible 
status 

1. Waiting for 
pressing HEAT-key 

 

Off 

last selected 
temperature 

last selected 
time-span 

Press: => 
Status 2 

 

Ignore 

Selection 
possible 

 

Off 

After pressing 
HEAT-key: 
Status 2 

 

 

2. Heating up 

 

 

On 

 

 

Flashing 

 

 

On 

 

Press => 
Status 1 

 

 

Ignore 

 

Selection 
possible 

Heating  up 
or  Cooling 
down 

If selected 
temperature is 
reached: 
Status  3 

 

 
 
 
 
 
 

3. Waiting for 
pressing START-key 

 

 
 
 
 
 
 
 

On 

 

 
 
 
 
 
 
 

On 

 

 
 
 
 
 
 
 

On 

 

 
 
 
 
 
 

Press => 
Status 1 

Press for t = 
30,  60,  120 

 

 
 
 
 
 
 

Selection 
possible 

 
 
 
 
 

Temperature- 
controlled  for 
stability  of 
the  selected 
temperature 

After pressing 
Start-key: 
Status 4 
(t



 
 
 

 

Ignore for 
t=



If selected 
temperature  will 
be  changed: 
Status  2 

If temperature 
to  different  to 
selected  one: 
Status  2 

4. Time: Count 
down 

 

On 

 

On 

 

Flashing 

Press => 
Status 1 

 

Ignore 

 

Ignore 

Stable 
Temperature 

End of count 
down: 
Status 1 

 

Heat-LED 

On: 

Heating up or stabilizing selected temperature. 

 

 

Temp.-LED 

Off: 

 

Flashing: 

Heater is  off 

 

Selected  temperature  is  not  reached  (heating  up  or  cooling  down) 

Together  with  HEAT-LED  On:    Selected  temperature  is  reached 
Together  with  HEAT-LED  Off:   Indicates  selected  temperature  without  indication  of  the  real  temperature  of  the  reactor 

 

Time-LED 

Flashing: 

Count down function 

 

On: 

indication of selected time-span without count down function. 

Содержание HydroTest HTCOD-TR

Страница 1: ...HTCOD TR HydroTest COD Thermoreactor Instruction Manual...

Страница 2: ...symbols 3 1 2 Important information 4 1 3 Unpacking 4 1 4 Connecting 4 1 5 Buttons 6 1 6 Function of buttons 6 2 Work session 19 3 Beeper 19 4 Function schematics 20 5 Maintenance 21 6 Cleaning 21 7...

Страница 3: ...and specified in the corresponding method descriptions Don t exceed temperatures or time spans in any case All warning labels must NOT be removed and should be replaced if they become damaged or fade...

Страница 4: ...the ventilation slots Air circulation is necessary to prevent the reactor from overheating If the air circulation is restricted it could cause fire or damage the reactor Never open the reactor housin...

Страница 5: ...he packing Part list Part Quantity 1 Thermoreactor HTCOD TR 1 2 Power cord European version 1 3 Instruction manual 1 1 4 Connecting On the reverse side of the reactor Selector for voltage 115 V 230 V...

Страница 6: ...matically switches off Heat key By pressing this key after switching the instrument on using the main on off switch see page 17 the reactor will heat up to the pre selected temperature Temp key By pre...

Страница 7: ...nently By pressing the Start key the timer starts the work cycles corresponding to the cho sen time span indication by Time LED Starting the work cycle the Time LED changes from lighted to flashing Wh...

Страница 8: ...aiting for pressing START key On On On Press Status 1 Press for t 30 60 120 Selection possible Temperature controlled for stability of the selected temperature After pressing Start key Status 4 t Igno...

Страница 9: ...nstructions set out in the Material Safety Data Sheet MSDS Chapter 6 and or 13 A contaminated aluminium block must be replaced prior to further use of the reactor The reactor should be sent to the man...

Страница 10: ...10 8 Wiring diagramm...

Страница 11: ...ff off on on yes 4 Wiring problem internal connection incomplete on off on off off yes 5 Heating problem no power reactor power probe problem on off on off on yes 6 T value underrange Mainboardfaulty...

Страница 12: ...12 Trace2o Limited The Technology Centre Wagtech Court Station Road Thatcham Berkshire RG19 4HZ Tel 44 0 1635 566772 Fax 44 0 1635 873509 sales Trace2o com www Trace2o com...

Отзывы: