Trace2O HydroTest HTCOD-TR Скачать руководство пользователя страница 11

 

 

9.

 

Err
or
 C

od
e

 (L

ED
 indic

a

tion)

 

11
 

 

 
 
 
 
 

No.

 

Error type

 

Possible  reason(s)

 

Temp LED

 

LED 30 min

 

LED 60 min

 

LED 120 min

 

LED 



Continuous  beep

 

 

 

 

 

Power  supply  frequency 

Frequency  higher/lower 
50Hz  /  60Hz; 
Mainboard  faulty 

 

 

on 

 

 

off 

 

 

off 

 

 

off 

 

 

on 

 

 

yes 

Safety  feature  reaction 

Mainboard faulty 

on 

off 

off 

on 

off 

yes 

ADC  error 

Mainboard faulty 

on 

off 

off 

on 

on 

yes 

 

 

Wiring problem 

internal  connection 
incomplete 

 

on 

 

off 

 

on 

 

off 

 

off 

 

yes 

 

 

 

 

Heating problem 

-

 

no power; 

-

 

reactor power 

-

 

probe problem 

 

 

on 

 

 

off 

 

 

on 

 

 

off 

 

 

on 

 

 

yes 

T  value  underrange 

Mainboard faulty 

on 

off 

on 

on 

off 

yes 

T  value  overrange 

Mainboard faulty 

on 

off 

on 

on 

on 

yes 

 

 

Temperature too high 

Probe  connection  faulty 
Mainboard  faulty 

 

on 

 

on 

 

off 

 

off 

 

off 

 

yes 

 

 

 

 

Microprocessor  failure 

 

 

EMC-interference 

100° on 
120° off 
150° on 

 

 

on 

 

 

off 

 

 

on 

 

 

off 

 

 

no 

 

10 

Temperature on 
mainboard too high 

 

Unit  overheated 

 

on 

 

on 

 

on 

 

off 

 

off 

 

yes 

Содержание HydroTest HTCOD-TR

Страница 1: ...HTCOD TR HydroTest COD Thermoreactor Instruction Manual...

Страница 2: ...symbols 3 1 2 Important information 4 1 3 Unpacking 4 1 4 Connecting 4 1 5 Buttons 6 1 6 Function of buttons 6 2 Work session 19 3 Beeper 19 4 Function schematics 20 5 Maintenance 21 6 Cleaning 21 7...

Страница 3: ...and specified in the corresponding method descriptions Don t exceed temperatures or time spans in any case All warning labels must NOT be removed and should be replaced if they become damaged or fade...

Страница 4: ...the ventilation slots Air circulation is necessary to prevent the reactor from overheating If the air circulation is restricted it could cause fire or damage the reactor Never open the reactor housin...

Страница 5: ...he packing Part list Part Quantity 1 Thermoreactor HTCOD TR 1 2 Power cord European version 1 3 Instruction manual 1 1 4 Connecting On the reverse side of the reactor Selector for voltage 115 V 230 V...

Страница 6: ...matically switches off Heat key By pressing this key after switching the instrument on using the main on off switch see page 17 the reactor will heat up to the pre selected temperature Temp key By pre...

Страница 7: ...nently By pressing the Start key the timer starts the work cycles corresponding to the cho sen time span indication by Time LED Starting the work cycle the Time LED changes from lighted to flashing Wh...

Страница 8: ...aiting for pressing START key On On On Press Status 1 Press for t 30 60 120 Selection possible Temperature controlled for stability of the selected temperature After pressing Start key Status 4 t Igno...

Страница 9: ...nstructions set out in the Material Safety Data Sheet MSDS Chapter 6 and or 13 A contaminated aluminium block must be replaced prior to further use of the reactor The reactor should be sent to the man...

Страница 10: ...10 8 Wiring diagramm...

Страница 11: ...ff off on on yes 4 Wiring problem internal connection incomplete on off on off off yes 5 Heating problem no power reactor power probe problem on off on off on yes 6 T value underrange Mainboardfaulty...

Страница 12: ...12 Trace2o Limited The Technology Centre Wagtech Court Station Road Thatcham Berkshire RG19 4HZ Tel 44 0 1635 566772 Fax 44 0 1635 873509 sales Trace2o com www Trace2o com...

Отзывы: