![Trace2O HydroTest HTCOD-TR Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/trace2o/hydrotest-htcod-tr/hydrotest-htcod-tr_instruction-manual_1146896011.webp)
9.
Err
or
C
od
e
(L
ED
indic
a
tion)
11
No.
Error type
Possible reason(s)
Temp LED
LED 30 min
LED 60 min
LED 120 min
LED
Continuous beep
1
Power supply frequency
Frequency higher/lower
50Hz / 60Hz;
Mainboard faulty
on
off
off
off
on
yes
2
Safety feature reaction
Mainboard faulty
on
off
off
on
off
yes
3
ADC error
Mainboard faulty
on
off
off
on
on
yes
4
Wiring problem
internal connection
incomplete
on
off
on
off
off
yes
5
Heating problem
-
no power;
-
reactor power
-
probe problem
on
off
on
off
on
yes
6
T value underrange
Mainboard faulty
on
off
on
on
off
yes
7
T value overrange
Mainboard faulty
on
off
on
on
on
yes
8
Temperature too high
Probe connection faulty
Mainboard faulty
on
on
off
off
off
yes
9
Microprocessor failure
EMC-interference
100° on
120° off
150° on
on
off
on
off
no
10
Temperature on
mainboard too high
Unit overheated
on
on
on
off
off
yes