Arsenometer HMAS100
Instruction Manual v2.0
2.3.2. PLATING
The plating step takes a few minutes.
There are two preliminary steps required for any analysis with the Arsenometer. The
plating step uses the AU500 Au Plating Solution and forms a plate on the working
electrode, covering the black tip with a yellow/ gold colour. The conditioning of the
electrode occurs when the element of interest is selected in the Test Methods menu
and then choosing condition electrode which is at the
top of the list when the element is
selected. The conditioning step should be carried out in sample water, and undertakes
a run which sensitises the Working Electrode.
Example:
Overview of As (III) procedure to first plate then sensitise the electrode:
Main Menu
Choose “Test Methods”
These plating steps
deposit the plate on
the tip of the
Working Electrode
Test Methods Menu
Choose “M2&3 Conditioning”
Half-fill SAB with AU500 Au Plating Solution
After plating, check tip for yellow / gold plate
Return AU500 Au Plating Solution to bottle
These
conditioning steps
(within the metal
menus) prepare
the Working
Electrode for
analysis.
Select analysis method...
Place buffer in SAB and fill SAB with
sample
From the
Test Methods Menu
, choose
As(III)
, and then
Condition Electrode
Refer to Application notes for Operating Procedures particular to each parameter.