Страница 1: ......
Страница 2: ...ical interference by the customer or by third parties with DewCheck Series 4 and related equipment without TQC s written permission No liability will be accepted for defects damage or injury caused du...
Страница 3: ...2 Log manual 4 2 3 3 Log interval 4 2 3 4 Setup 4 3 Setup 5 3 1 Interval 5 3 2 Limits 6 3 3 Language 7 3 4 C F 7 3 5 Date time 7 3 6 Backlight on 7 3 7 Device info 8 4 Measuring 9 4 1 The Measure func...
Страница 4: ...k 21 5 1 1 Positioning the DewCheck4 with Dock 22 5 2 DewLog V1 the software 23 5 2 1 Installation 23 5 2 2 Quick start 23 6 Safety Care and Maintenance 25 6 1 General Maintenance 25 6 2 Battery maint...
Страница 5: ...d at regular intervals It is not only important to take frequent measurements they should also be taken at different areas of the surface Factors such as the sun shadow side and the shape and the mass...
Страница 6: ...re described below Press the ON OFF button to switch the instrument on and off Press to navigate upward through the menu Press to navigate downward through the menu Press to access the selected functi...
Страница 7: ...the surface temperature An external sensor can be used in combination with the optional DewCheck4 Dock See Section 4 4 for detailed information on parameters 2 3 The Start menu When DewCheck4 is swit...
Страница 8: ...rt measuring in a batch and manually log the measurements See Section 4 2 for more information 2 3 3 Log interval Select Log interval to start measuring in a batch and log the measurements in a set in...
Страница 9: ...acklight on Device info These submenus are described below in detail 3 1 Interval Select Interval to set the standard interval time for the measuring option Log interval Each interval time is paired w...
Страница 10: ...nge the batteries or change the interval time to be able to complete interval logging 3 2 Limits The limits you set here are only used when measuring without logging option Measure in the start menu T...
Страница 11: ...te time Use this option to set the date and time The date and time are displayed on the start menu Date Select the date format and enter the correct date in the chosen format The selected format is di...
Страница 12: ...ry status full and the memory status free See Section 6 2 for more information on battery maintenance Figure 3 4 Device info example The battery status is also displayed on the measurement screens The...
Страница 13: ...mended measuring method See Section 4 5 for more information on batches 4 1 The Measure function Select Measure in the start menu to start measuring immediately without logging the results If required...
Страница 14: ...dicator between the parameter and the reading indicates if the value is increasing or decreasing When a limit is exceeded an audible alarm is triggered and the alarm sign displayed next to the concern...
Страница 15: ...PC software You can select empty batches and batches that were previously used for manual logging Closed batches cannot be used for manual logging 3 Press OK The following screen displays the availabl...
Страница 16: ...ensor against the surface at an angle of 90 see figure 4 6 and keep it in this position until the measured value remains stable DO NOT USE EXCESSIVE FORCE A light contact with the surface is sufficien...
Страница 17: ...oring it Use Cancel to cancel the measurement and prevent it being stored 9 Repeat steps 7 and 8 to save more log records 10 Select Back to stop manual logging and return to the Batch menu See Section...
Страница 18: ...u can start logging Figure 4 8 Batch selection screen 3 Select a Batch Batches can be numbered of given names using the PC software Only empty batches can be selected for interval logging 4 Press OK T...
Страница 19: ...f 90 see figure 4 11 and keep it in this position until the measured value remains stable DO NOT USE EXCESSIVE FORCE A light contact with the surface is sufficient Using force does not increase the me...
Страница 20: ...cerned parameter 8 Press OK to start interval logging The Start option is replaced by the Stop option The current log record and interval time are displayed at the bottom of the screen Each log record...
Страница 21: ...e same way see Figure 4 13 Use the Setup menu to set the limits for each parameter when measuring without logging see Section 3 2 Use the Batch menu to set the limits for each parameter when measuring...
Страница 22: ...an angle of 90 see figure 4 14 and kept in this position until the measured value remains stable A light contact with the surface is sufficient DO NOT USE EXCESSIVE FORCE Using force does not increase...
Страница 23: ...m symbol displays as long as Delta T is lower than the set limit It is risky to paint in such a situation as the slightest change in climate conditions may cause condensation on the surface 4 5 Batche...
Страница 24: ...selected batch The Recall menu displays the Average Maximum and Minimum values for Delta T measurements in the saved dataset Clear The Clear function deletes all stored batches from memory It is not p...
Страница 25: ...screen or with the optional DewCheck4 Dock The stored data can be transferred to a personal computer using a USB cable 5 1 Installing the DewCheck4 Dock 1 Connect the probe to the Dock making sure th...
Страница 26: ...1 Positioning the DewCheck4 with Dock In its role as magnetic holder the Dock can be fixed to any ferrous steel surface for a longer period of time The Dock must then be secured with a bolt Figure 5...
Страница 27: ...utton and open the Product key txt file Copy the key into the text box and click Install Follow the instructions of the Install Shield Wizard which will guide you through the installation process 5 2...
Страница 28: ...d Display units and Synchronize the date and time with the PC Step 9 Click Next Step 10 Set the preferred Interval time Step 11 Click Next Step 12 Enter the batch names Step 13 Click Next Step 14 Enab...
Страница 29: ...it may be damaged This can happen if it is left in a car parked in strong sunlight We suggest cleaning the instrument with a soft dry cloth such as those made to clean eyeglasses Do not use solvents...
Страница 30: ...attery symbol flashes The battery symbol is displayed in the measurement screens The battery status is displayed in the Setup menu submenu Device info To save the time and date settings the batteries...
Страница 31: ...20 80 C 4 175 F Temperature Ts Surface Accuracy 0 5 C 1 F Measurement resolution 0 1 C 0 1 F Measurement range 30 60 C 22 140 F Temperature Te Surface External Accuracy 0 5 C 1 F Measurement resoluti...
Страница 32: ...ment features Limits Adjustable for each parameter Lo Hi Alarms Beep sound symbol in display Hold freeze function Yes Data storage Two types Manual and interval auto Data recall Gauge displays average...
Страница 33: ...l logging 1 year with 10 min interval measurement manual logging 400 hours continuous operation data based upon use without backlight Optional DewCheck4 Dock Unit to be connected to the back of the ga...
Страница 34: ...time are displayed at the bottom of the main menu 8 2 Measure menu Measure Hold Release Menu 8 3 Log manual menu Log Manual Batch 1 Batch 2 Batch 3 etc up to Batch 8 Batch 1 Log Hold Back Recall Avera...
Страница 35: ...Log interval menu Log Interval Batch 1 Batch 2 Batch 3 etc up to Batch 8 Batch 1 Log Hold Back Recall Average Maximum Minimum Back Limits Limits On Limits Off Set Limits Back Clear Yes No If a Batch...
Страница 36: ...me Back Backlight On Off Trademark Copyright TQC Thermimport Quality Control B V Nijverheidscentrum 14 2761JP Zevenhuizen The Netherlands E mail info tqc eu Internet site http www tqc eu This manual m...