| 24
1.
Select Scan Setup in the Main Menu and press OK.
2.
Select Calibrate and press OK. The Calibrate menu is displayed.
3.
Select Values and press OK to display a list of calibration values for each scan angle.
4.
Use the arrow buttons to select the scan angle for which you want to adjust the standard calibration
values and press OK. The scan angle and the current standard value are displayed.
5.
Use the Down arrow to select the scan angle and press OK.
6.
Use the arrow buttons to adjust the calibration value:
•
Press the arrow buttons briefly to increase or decrease the value in steps of 0.1.
•
Keep the arrow buttons pressed down to increase or decrease the value in steps of 5.
7.
Once the correct calibration value has been reached, click OK.
Note: The calibration values are printed at the bottom of the calibration holder next to the glass plate.