17.2 Turbidity Troubleshooting
Symptom
Possible Causes
Remedy
Turbidity cannot
be calibrated, or
readings
inaccurate
1. Turbidity standard is
incorrect.
2. Turbidity sensor is dirty.
3. Turbidity sensor is not
immersed correctly.
Check Turbidity standards.
Zero calibration must be
performed in fresh Distilled
water. Check that the
Turbidity standard has been
correctly entered (section
5.1).
Clean tip of Turbidity sensor
with a clean, soft cloth. Avoid
scratching the sensing
surface.
Turbidity sensor must be at
least 30mm above the floor
of the sample vessel.
Meter displays
“
No Probe
” when
the sensor is
connected.
1. Faulty cable
2. Faulty sensor
Repair or replace the
detachable cable.
Repair or replace the sensor.