Page 5
¬
Alarm LED
The Alarm LED provides visual indication of when the control limit has been
exceeded. The control limit can be set for a “too high” or a “too low” alarm.
See section 3.5.
-
Limit Control
Adjust this control to the Salinity/TDS value at which the Alarm LED and
(optional) relay output are activated. See section 3.5.
®
Zero Calibrate Control
Adjust this control to Zero calibrate the
mini
CHEM
-Sal
with the sensor dry
and in air. See section 4.1.
¯
Span Calibrate Control
Adjust this control to Span calibrate the
mini
CHEM
-Sal
in a Salinity/TDS
standard. See section 4.1.
°
Function Switch
The Function Switch is used to select the mode of operation. See section 2.
±
Display Window
The
mini
CHEM
-Sal
has a large, easy to read LCD display. The units of
measurement are clearly shown in the window beneath the display.
Æ
Terminal Cover
The terminal cover provides easy access to all of the connections, (optional)
relay output fuse, configuration jumpers and user-adjustable trimmers. The
cover is water resistant to IP65.