10. When the reading has stabilised, adjust the
SPAN
control until the display
shows the value of the standard.
11. Rinse the Salinity/TDS sensor in distilled water and blot dry.
12. The
mini
CHEM
-Sal
is now calibrated and ready for Salinity/TDS
measurements.
4.2 Calibration Notes
1. The Zero calibration is quite stable long term, and only needs to be performed
monthly as a routine check. In applications where the sensor can become dirty
or coated with oils etc., a Zero calibration may need to be done weekly.
2. A Span calibration should be performed at least weekly. Of course, more
frequent calibration will result in greater confidence in results.